Analysis of Corneal Response to Intra-Stromal Ring-Segment Implantation in Human Eyes. A Pathology Study

J. Feito

1A

, A. Astudillo

2, 1A

, B.Baamonde

2, 1B

, J. Alfonso

3, 2

, J Merayo-Lloves

3

.

1

Hospital Central de Asturias, Oviedo, Spain;

A

Pathology;

B

Ophthalmology

2

Surgery, University of Oviedo, Oviedo, Spain

3

Fundación Investigación Oftalmológica & Instituto Fernández-Vega, Oviedo, Spain

INTRODUCTION

Intracorneal ring (ICR, consisting of one segment) or

intracorneal ring segments (ICRS, consisting of two semi-rings) refer to plastic ring (made from Poly-methyl metacrylate,

PMMA) arcs inserted into the corneal stroma. ICR⁄ICRS can be used for: correction of low grades of myopia, treatment of

keratoconus sometimes as an (alternative to transplantation) and for correction of residual myopia after laser treatment of severe myopia. An important advantage of ICRS is that the

operation is reversible, i.e. the ring can be removed or replaced if necessary as only affects the periphery of the cornea.

We want to evaluate the pathology findings observed in a

patient who had an ICRS ring inserted on his cornea and who afterwards had penetrating keratoplasty.

MATERIALS & METHODS

Human cornea implanted with PMMA ring-segments (Ferrara ring, Ferrara & Hijos SL, Boecillo, Spain) from patient with keratoconus were analyzed for pathology studies: conventional histology and immunohistochemistry. The ICRS was placed using mechanical dissection and retired because there was a risk of corneal perforation after a temporal extrusion of the segment.

Keratoplasty was performed 5 months later.

The corneal button was bisected for histological and immunohistochemical study. Tissue sections were placed on formalin-fixed paraffin-embedded blocks stained with Hematoxilin-Eosin, Masson trichrome, and periodic acid-Schiff solutions and examined by light microscopy.

Immunohistochemistry was performed on the sections of

hemicorneal button. Tissue sections were dewaxed in xylene, rehydrated through a graded ethanol series, and washed with phosphate-buffered saline. Antigen retrieval was achieved by

heat treatment in a water bath (20 minutes at 96 ºC) using a pH 9 target retrieval solution. Samples were stained using a sensitive polymer-based detection system (En Vision Plus,

DakoCytomation automated immunostainer). The reference, and dilution factor of primary antibodies were as follows: vimentin

(Biogenex MU074-UC, 1/200), and CD34 (Dako M7165, 1/100);

to detect keratocytes; a marker of stromal cell activation CD 10 (Novocastra NCL-L-CD10-270, 1/80); a marker of marophagic phenotype CD 68 (Dako M0814, 1/2000); and a-smooth muscle actin to detect myofibroblast phenotype (Dako M851, 1/50).

RESULTS

We observed an atrophic epithelium in the area above the implant with a decrease in the number of layers and thinning of the few layers remaining, No irregularities or disruptions were observed in the Bowman membrane. There was also a reduction in the number of keratocites over the implant.

Around the implant there was a de-orientation of the cells and hypercelularity, staining positive for SMA that mark the stromal cells as

differentiated towards the myofibroblast phenotype. The stroma around the ICR segments stained more intensely with PAS than normal corneal stroma do. Also, adjacent to the implants there were stromal cells positive for CD 10 (neutral endoprotease), CD 34 (keratocites). Interestingly, there were keratocites positive for CD 68 surrounding the channel (macrophagic phenotype).

These findings suggest that ring segments implants modulate the epithelial and corneal stroma with atrophy of the epithelium and adjacent stroma and changes in the phenotype of corneal keratocites around the ring segment.

CONCLUSIONS

Although these PMMA corneal inserts

appear to be well tolerated, from the results of long-term clinical studies in humans, has been demonstarred the changes in

keratocyte phenotype suggesting an active tissue repair process and physiological

stress.

Intra-stromal ring segments reduce the

number of keratocites and epithelial cells and modulate the phenotype of keratocites with expression of markers for

macrophages (CD 68), myofibroblast (SMA) and endoprotease (CD10).

None of the authors have commercial interest on this study

REFERENCES

1.- Histopathological Analysis of Post-Laser-Assisted In Situ Keratomileusis Corneal Ectasia With Intrastromal Ring Segments. Spirn MJ, Dawson DG, Rubinfeld RS, Burris C, Talamo J, Edelhauser HF, Grossniklaus HE. Arch Ophthalmol. 2005 Nov.

2.- Histopatholocial findings after intracorneal ring segment implantation in keratoconic human corneas.

Samimi S, Leger F, Touboul D, Colin J. J Cataract Refract Surg. 2007 Feb.

3.- Histologic Evaluation of Corneal Stroma in Rabbits After Intrastromal Corneal Ring Implantation. Twa MD, Ruckhofer J, Kash RL, Costello M, Schanzlin DJ. Cornea. 2003 Mar.

D901¿?

Hematoxilin-Eosin 10x Smooth Muscle Actin 20x

CD 10 10x

DISCUSSION

The findings of epithelial atrophia and the disminution in the number of keratocytes above the implant are already well

described. We haven´t observed changes in the Bowman membrane although they have been related with keratoconus. There is

also evidence about a immunophenotipic change in the keratocites. The SMA

staining is already described, also the

reduction in the CD 34 positive keratocites.

To our knowledge, this is the first report about the CD 10 and CD 68 role in the

apoptotic changes tipical of this implanted corneas.

Vimentin 20x

PAS 10x CD 68 10x

Referencias

Documento similar

In our approach, we have defined and analyzed the following variables: the areas of the anterior and posterior corneal surfaces of the solid model generated (A ant , A post ), the

Q45 and Q8, corneal asphericity in the central 4.5 and 8 mm; RMS, root mean square; HOA, high order aberrations; SA, spherical aberration; ΔVOL mct , change in corneal volume

The purpose of the current study was to determine the geometrical profile and novel correlations of the corneal deformation for KC eyes with mild visual limitation compared to

Also, some significant positive correlations between corneal volume and sagittal plane apex area of the posterior surface (R 2  0.919, p ≤ 0.001), and with sagittal plane area

The results ob- tained in this work suggest that the best predictive biometric parameters are an- terior corneal surface area and posterior apex deviation, and that the

points used hereafter [3] (Figure 2): A ant is the area of the anterior corneal surface [1,2,4]; A post is the area of the posterior corneal surface [1,2,4]; CV is the total

• VOLmct: volume contained in the intersection between the solid model of the cornea and a cylinder of revolution with radius × (from 0.1 to 1.5 mm) and its axis defined by the

The purpose of the current study was to determine the geometrical profile and novel correlations of the corneal deformation for KC eyes with mild visual limitation compared to