Developing a prototype of pCT scanner

J.A. Briz

a

, E. Nácher

b

, I. Posadillo

a

, V.G. Távora

a

, O. Tengblad

a

, M.J.G. Borge

a

, A. Ortiz

a

, A. Perea

a

, J. Díaz-Ovejas

a

, S. Viñals

a

aInstituto de Estructura de la Materia, CSIC, Madrid, Spain

bInstituto de Física Corpuscular, CSIC-Universidad de Valencia, Valencia, Spain

Corresponding autor: [email protected]

Developing a prototype of pCT scanner – III Jornadas IFIMED/RSEF 2020 1

UCM-CIEMAT-IEM(CSIC)

(2)

Proton vs. X-ray therapy

Main advantage with respect to conventional radiotherapy (X-rays):

Less damage produced in surrounding healthy tissues -Almost no dose deposited beyond the tumoral region -Lower dose in region before the tumor

Protons X-rays

Proton vs IMRT (X-rays) treatment dose plans for a 46 years old

woman with carcinoma

(3)

The importance of proton-CT

Proton-therapy treatment plans are based on X-rays CT images (XCT)

• Maps of Hounsfield Units (XCT) à RSP (Proton-therapy)

(relative stopping power)

( μ à attenuation coefficients)

• Conversion induces uncertainties in proton ranges of ~3%

Using proton-CT (pCT) images:

uncertainties can be reduced to < 1%

Challenge: design a pCT scanner providing high quality images

Developing a prototype of pCT scanner – III Jornadas IFIMED/RSEF 2020

¡¡ Absorbed dose for

-radiograph: 0.01 mGy -X-ray CT: 30-50 mGy -proton CT: ~1.3-1.4 mGy

Robert P. Johnson et al.

Rep. Prog. Phys. 81 (2018) 016701 3 Conversion

(4)

IEM-CSIC prototype of pCT scanner

Area: 50 x 50 mm

2

Segmentation: 256 3x3 mm

2

pixels

• Phoswich: LaBr

3

(4 cm) + LaCl

3

(6 cm) in Array of four 27x 27mm

2

• Energy resolution:

• 3 % for 1173 keV g (

60

Co)

• 3-5 % for 60 – 130 MeV protons

• decay time 16 ns

O. Tengblad et al. NIM A 704 (2013) 19-26

CALIFA Endcap Phoswich Array (CEPA4)

pCT scanner

(DSSD)

(5)

Proof-of-concept: CMAM Experiment

• Centro de MicroAnálisis de Materiales (UAM)

• 10 MeV protons à thin “phantoms” ~1 mm

• Beam intensity ~1 nA

• 2 DSSDs detectors:

§ 60 µm

§ 500 µm

Developing a prototype of pCT scanner – III Jornadas IFIMED/RSEF 2020 5

Primary Beam Scatte red Be am

24

o

Detectors and phantom

Phantom

Detector1

Detector2

Scattering target

(6)

“Phantoms”

2 Master theses defended in Univ. Sevilla in Sept 2019

Experimental results

Hole

1 mm Al

200 µm Al + 500 µm PVC

200 µm Al 500 µm PVC with

Holes of 4, 6 y 8 mm 500 µm PVC

+ 200 µm Al

Geant4 simulations

Proof-of-concept: CMAM Experiment

(7)

Future experiments

Developing a prototype of pCT scanner – III Jornadas IFIMED/RSEF 2020

Beam line Scat. target

Exp. chamber Scattered beam proton

beam

Experiments to be performed in 2021 at:

1º- KVI-CART (Groningen, Netherlands) à 2º- CCB (Krakow, Poland) à

Higher beam energies: 50-230 MeV

Larger phantoms: from 60 mm to 200 mm thickness

7

Scattered Proton beam 50-230 MeV

Objectives

• Properties of direct images

(radiographs) (Single projections)

• Tomography of 3D objects

(Multiple projections):

o test of reconstruct. algorithms o evaluation of image properties

Scheduled For March Approved 2021

and to be

scheduled

(8)

Phantoms

Single projection Multiple projections

• PMMA

• Aluminum • Air

• Water • Alcohol

Simple (2 inserts)

Alcohol Water

Derenzo

Top view

Spatial resolution

• Sensitivity different materials

• Imaging spatial shapes

Reconstructed CT images

Simple (Cross) Derenzo-type

Aluminum inserts

• Sensitivity different materials

• Imaging spatial shapes Spatial resolution

Radiograph mode

Materials

(9)

pCT scans: Geant4 simulations

Developing a prototype of pCT scanner – III Jornadas IFIMED/RSEF 2020

Scattered proton beam

DSSD1 Phantom DSSD2 CEPA4 Rotations

Geant4 geometry Image obtained at 0º 37 projections

in 5º steps from 0 to 180º

Reconstructed Image using Filtered-Back

Projection algorithm (FBP)

100 MeV proton beam

Extensive studies with simulations done and ongoing to optimize our setup

Reconstruction algorithms

Alcohol Water Water Alcohol

9

(10)

Preparation of Experiment

Current status of experimental setup:

Experiments to be performed

(if COVID-19 situation let it possible)

-KVI-CART scheduled for March 2021 -CCB Krakow: to be scheduled

Beam

Experimental tests ongoing at

IEM-CSIC Lab to use new digital

electronics and test detectors.

(11)

Summary

• Successful proof-of-concept experiment performed at CMAM (June 2019)

• 2 future experiments approved

-Proton beams from 50-230 MeV

-Phantoms from 6 cm thick to 15-20 cm

-Properties of single projections (radiographs)

-Properties of CT scans: multiple angle projections

-Test different reconstruction algorithms for CT reconstruction

• Monte Carlo simulations to optimize experimental setups

• Ongoing experimental tests in IEM-CSIC Lab

Developing a prototype of pCT scanner – III Jornadas IFIMED/RSEF 2020 11

KVI-CART (Groningen, Netherlands) CCB (Krakow, Poland)

Scheduled for March

2021

Approved

and to be

scheduled

(12)

Thanks for your attention