3.2.2.1 Overview

The preferred reference method for measurement of arsenic in air, outlined further below, involves chemical analysis of filters collected using the high volume method for gravimetric determination of PM10, following United States Environmental Protection Agency (US EPA) protocols (MfE, 2002). Applied to arsenic, this method assumes that a) almost all arsenic in air is likely to be in particulate (rather than vapour) form, and b) this arsenic will be present in the PM10 fraction (Maggs, 2000). As the fraction retained by the lungs, PM10 should also represent the best estimate of inhalation exposure to arsenic in air.

3.2.2.2 Sampling frequency and rationale

Typically, reference methods used for monitoring concentrations of metals in

particulate matter by the gravimetric method rely on a non-automatic manual ‘batch’ sampling approach following US EPA protocols. Regulatory monitoring requirements usually adopt a consecutive 1-in-3 day or 1-in-6 day sampling regime which is deemed sufficient to calculate a representative annual average from a long term sequence of low-time resolution data (Bortnick & Stetzer, 2002). The non-consecutive sampling frequency also reflects the resource-intensive nature of non-automatic sampling that requires manual filter changes and off-site laboratory preparation and analysis. This sampling strategy may not be appropriate in situations where data variance is high (e.g., due to strong seasonality) or where health effect assessment from short term exposure periods (e.g., winter only or 24-hour) is required.

Previous elemental speciation of particulate matter at both Raumati South (Kapiti Coast) and Wainuiomata, which was undertaken for receptor modelling (source apportionment) indicates that arsenic concentrations are very low or non-detectable during the non-winter period and are elevated and highly variable during the winter months when people use fires for home heating. Therefore, in order to account for the strong seasonal variability in arsenic concentration, sampling frequency was increased from 1-in-3 day to 1-in-2-day during the winter months (May to August inclusive) over the two year monitoring period. As a result of this sampling strategy, the population

estimate of the mean will be more precise as the more variable strata (winter) was sampled more intensively than the less variable strata (non-winter). Section 3.3.4 discusses the method for taking into account the difference in sampling frequency when calculating the annual means.

Sampling was undertaken from 22 October 2011 until 29 October 2013, providing a 24- month period that captures two winters and enables inter-annual variation to be assessed.

3.2.2.3 PM10 gravimetric sampling

PM10 was sampled using a high volume sampler (Lear Siegler, serial no. A107) fitted with a PM10 size selective inlet (Graseby Andersen, serial no. 4491). This instrument was operated in accordance with AS/NZS 3580.9.6:2003 Method 9.6: Determination of suspended particulate matter – PM10 high volume sampler with size selective method –

gravimetric method11. The instrument was calibrated monthly using a critical orifice calibrator (Lear Siegler, serial no. 026) to ensure flow rate was maintained between 65 and 75 m3 per hour as required by AS/NZS 3580.9.6:2003.

Sequentially numbered 8x10 inch Whatman grade QM-A quartz filters were conditioned and weighed at constant humidity by GNS Science in accordance with AS/NZS 3580.9.6:2003. These filters were exposed for 24-hours (midnight to midnight) on a one-in-three (1:3) day sampling regime which was increased to a one-in-two (1:2) day frequency between May and August. After exposure, filters were re-conditioned and re-weighed and then stored at -18oC to minimise loss of the volatile fraction of particulate matter prior to further analysis.

During the monitoring campaign, 14 field blanks were deployed for quality assurance purposes to ensure that filter handling and transport did not bias results. Field blanks are filters that were subject to the same handling procedures as sample filters except that the high volume sampler was not turned on and the filter was mounted in the sampler and then removed immediately. Likewise, laboratory blanks are filters that were prepared for use but were not removed from the laboratory environment to

11 This standard is based on US EPA 40 CFR Part 50, Appendix J – Reference Method for the Determination of Particulate Matter as PM10 in the

Atmosphere, which is the monitoring method recommended by the ambient air quality guidelines (Ministry for the Environment, 2002).

ensure that laboratory conditions were not biasing results. In total, 10 laboratory blanks were processed. According to AS/NZS 3580.9.:2003, field and laboratory blanks should be used at a frequency of not less than 1 in 20 (5%), which in this case would equal 14 filters.

PM10 concentrations per filter were calculated using Equation 3.1 corrected to standard temperature and pressure (STP), i.e., 273.15 K (0oC) and 101.3 kPa, using barometric pressure (hkPa) and temperature (oC) measurements from 2 m above ground level at the GWRC weather station at Wainuiomata Bowling Club. Volumetric flow rate was maintained at 70 m3 per hour based on the instrument’s flow rate set point. Equation 3.1: 𝐶𝑃𝑀10 =�𝑚𝑓 − 𝑚𝑖�× 10 3 𝑉 Where: CPM10 = concentration of PM10 (µg/m3) mf = final mass of filter (mg)

mi = initial mass of filter (mg)

V = volume corrected to STP, calculated as:

𝑉 = 𝑄 ×𝑡 × 273 𝑇

1 ×

𝑃1

101.3

Where:

V = volume of air sampled (m3) corrected to reference conditions of 0oC and 101.3 kPa

Q = volumetric flow rate (m3/minute)

t = sampling time (minutes)

T1 = mean ambient temperature over sampling period (K)

3.2.2.4 PM10 and PM2.5 continuous sampling

In addition to gravimetric 24-hour PM10 measurements, PM10 concentration was also measured by GWRC using a Thermo Scientific series FH62 C14 Beta Attenuation Monitor (BAM). This method is an US EPA Automated Equivalent Method for PM10 monitoring (EQPM-1102-150). The sample inlet heater was set to 40°C, in accordance with recommended best practice (MfE, 2009). PM2.5 concentrations were also

measured with a BAM fitted with a PM2.5 Very Sharp Cut Cyclone (VSCC) after the PM10 head. This method is not US EPA equivalent because the BAM is an older model

without a dynamic heating system. The filter tape was set to advance at midnight for both instruments, which were also both set to correct PM mass concentrations to 0°C and standard pressure.

3.2.2.5 Determination of arsenic concentration in PM10

Four subsamples consisting of a 47 mm diameter punch was taken by the author from each exposed high volume PM10 filter at the GNS Science laboratory facilities (Figure 3.2). Punches were taken using a stainless steel machine-milled cylinder and the cutting edge of the punch was cleaned with laboratory grade ethyl alcohol between each filter. A total of 303 filter punches (including 14 field blanks and 10 lab blanks) were couriered to Hill Laboratories (Hamilton) in a chilly bin in two batches. The first batch was sent for analysis 6 December 2012 (144 filters) and the second batch was sent on 16 August 2014 (159 filters).

The reference method (US EPA) for determining arsenic concentrations in PM10 is acid digestion followed by Atomic Absorption Spectrometry (AAS) or an equivalent method. The equivalent method used by Hill Laboratories was ICP-MS (Inductively Coupled Plasma Mass Spectrometry) which is faster, more precise and more sensitive than AAS. Hill Laboratories used a modified aquaregia digestion NIOSH method 7303, issue 1. The concentration of arsenic in air was calculated as per Equation 3.2.

Equation 3.2:

𝐶𝐴𝑠 = 𝐶𝑠𝑉 × 1000 ×𝑝

Where:

CAs = arsenic concentration in air (ng/m3) at STP

Cs = arsenic concentration in test solution (µg/sample)

V = volume of air sampled (m3) corrected to reference conditions of 0oC and 101.3 kPa

p = proportion of filter used (unitless), calculated as: 𝑝 = 𝑇𝑜𝑡𝑎𝑙𝐴𝑟𝑒𝑎𝑎𝑟𝑒𝑎𝑓𝑖𝑙𝑡𝑒𝑟𝑒𝑥𝑝𝑜𝑠𝑒𝑑𝑝𝑢𝑛𝑐ℎ𝑓𝑖𝑙𝑡𝑒𝑟

Where:

Filter punch = 17.33 cm2

Exposed filter = 419 cm2

The limit of detection (LOD) reported by Hill Laboratories for arsenic obtained by ICP-

MS is 0.05 μg per sample. Therefore the LOD for arsenic in ambient air (ng/m3) may differ slightly for each exposed filter depending on the volume of air sampled in that 24-hour period. In this case the LODs for arsenic concentrations ranged from 0.74 to 0.78 ng/m3.

3.2.2.6 Determination of black carbon and multi-element concentrations

Black carbon is the solid fraction of particulate matter that absorbs light strongly. Black carbon measurements (taken from one of the quadruplicate set of PM10 filter punches) were obtained by GNS Science using a light reflectance method with a M43D Digital

Smoke Stain Reflectometer (Ancelet, Davy, Trompetter, Markwitz, & Weatherburn, 2011a).

Multi-element concentrations in PM10 (using one of the quadruplicate set of PM10 filter punches) were determined using X-Ray Fluorescence spectroscopy (XRF) by GNS Science at the New Zealand Ion Beam Analysis Facility. A general description of the XRF method used for ambient air filters is provided in Davy and Ancelet (2014).