• No se han encontrado resultados

4. Resultados

4.2. Concentración de radionucleidos naturales y elementos estables en el trigo,

4.2.2. Plántulas

known, of course the measured values can be related to the exact refractive index and particle density, using the Mie scattering program and the geometry parameters of the KS-93 optics.

5.2.4 OPC counting efficiency

As the intensity of the scattered light decreases with decreasing particle size, not all small particles passing the optics are counted. The counting efficiency ( ) dp for a certain particle size dp is the ratio of the number of particles the OPC unit counts to the number of available particles. The latter one was measured with a reference instrument, in this study a CPC, type “TSI 3010” (Fig. 5.4).

Figure 5.7: Influence of the aerosol composition (complex refractive index, particle density) on the particle size distribution (a) and the particle mass concentration (b), measured with the CARIBIC OPC unit along a measurement flight between Frankfurt/Germany and Johannesburg/South Africa on November 15th, 2010. The analyzed data were related to an internal mixture of mid-latitude lowermost stratospheric aerosol (LSA) (green), mid-latitude upper tropospheric aerosol (UTA) (blue), and tropical mid tropospheric aerosol (MTA) (red), using a Mie scattering program. The relative difference from the UTA mass concentration is displayed in (c).

100 1000

relative difference from mid-latitude UT aerosol mass [%] tropical mid-tropospheric aerosol lower stratospheric aerosol

00:00 01:00 02:00 03:00 04:00 05:00 06:00 07:00 08:00 09:00

time [UTC]

00:00 01:00 02:00 03:00 04:00 05:00 06:00 07:00 08:00 09:00

time [UTC]

5.2 Calibration of the OPC unit

( )p OPC

CPC

d N

  N (5.8)

NOPC particle concentration measured with OPC [1/cm3] NCPC particle concentration measured with CPC [1/cm3]

The measured particle concentrations can be directly compared for each particle size when using the monodisperse latex calibration standard. However, as described in the discussion of Fig. 4.5 and Fig. 5.3, for particles larger than ~250 nm, erroneous counts occur at the small end of the size distribution. Consequently only the counts of the main peak were included in the calculation of the particle concentration.

To obtain more than four measurement points (latex calibration standards 125 nm, 200 nm, 350 nm, and 500 nm), AS was used too. Multiple charged particles were resolved in size and number by setting the OPC size distribution to 256 channels.

In this way the counting efficiency of the OPC unit was calculated gradually from large particle sizes to smaller ones. For each singlet maximum diameter the number of all multiple charged particles was summed up (Nmultiple). Thereby the counting efficiency for the doublets, triplets and larger particle sizes was taken into account (Eq. 5.9). The respective counting efficiencies were taken from the previous calibration steps. By subtracting Nmultiple from the number of particles the CPC counted (NCPC), the total number of singly charged particles was calculated (NCPC s, ; Eq. 5.10). Dividing the number of singly charged particles the OPC unit counted (NOPC s, ; Eq. 5.11) byNCPC s, , the OPC counting efficiency ( ( ) dp ) at the dp of the singlets was finally determined (Eq. 5.12). The obtained ( ) dp was used for the next smaller particle size calibration step. For the first calibration step (largest particles, 500 nm AS, 582 nm UTA), particles larger than the singlet size were not present in the OPC size distribution18. The absence of multiple charged particles is reasonable, because the used particle generator did not generate such large particles [TSI, 2008], hence Nmultiple = 0.

1

61

Ni number of counts in a certain size channel [1]

i counting efficiency for a certain size channel [1]

nOPC number of size channels in the OPC particle size distribution [1]; (nOPC= 256 for calibration) a first channel, in which the singlets were seen [1]

b last channel, in which the singlets were seen [1]

The resulting OPC counting efficiency curve is displayed in Fig. 5.8. Like for the size calibration, all particle diameters were related to UTA. Above about 250 nm, the OPC counting efficiency is more or less constant (maximum asymptotic counting efficiency).

For smaller particle diameters the counting efficiency decreases. The particle diameter where the counting efficiency is 50% of the maximum asymptotic counting efficiency

Figure 5.8: OPC unit counting efficiency for mid-latitude UT aerosol particles (mc,total(λ) = 1.48-0.0143·i (λ = 830 nm)). Black squares represent measurement, where only the main peaks of the singlets were taken into account. Red triangles represent calculations where the erroneous counts (cf. Sec. 5.1) were included. The x-axis error bars indicate the maximum uncertainty due to the combination of the different aerosol type calibrations and the Gaussian width of the DMA distribution (Sec. 5.3.3). The y-axis error bars indicate the maximum uncertainty concerning the measurement air volume (Sec. 5.3.3). As the black and the red measurement points were obtained from the same dataset, a potential shift of the true counting efficiency (within the error bars) would be in the same direction for both curves. Hence, the overlap of the positive error bar (peak counts) with the negative error bar (all counts) does not indicate that the counting efficiency might be equal.

100 200 300 400 500 600 700 800

0.0

5.2 Calibration of the OPC unit

(dp,50), is a characteristic size for the lower detection limit of an optical aerosol counting instrument. Using Eq. 5.13 as fitting function [Wiedensohler et al., 1997], the UTA dp,50 was calculated to be 140.5 nm. For the LSA calibration dp,50 is 146 nm and 141 nm for the MTA, respectively.

1

dp particle diameter [nm]

A1 approximation parameter [1]

A2 maximum asymptotic counting efficiency [1]

,1

dp approximation parameter [nm]

,1

dp approximation parameter [nm]

In addition to the true counting efficiency (black squares), Fig. 5.8 illustrates the effect of considering the erroneous counts too (red triangles). For particle sizes smaller than

~250 nm only few erroneous counts occurred, influencing ( ) dp only slightly (cf. also discussion to Fig. 5.3). For particle measurements larger than about 250 nm, ( ) dp increased significantly above 100%. Because for all measurements the used particle size was much larger than the CPC dp,50(~10 nm), the CPC was assumed to count 100%

[TSI 2002]. Consequently, OPC counting efficiencies much higher than 100% are unrealistic. Hence, the additional small pulses must be erroneous. As the black and the red measurement points were derived from the same dataset, the overlap in the error bars (positive for peak counts vs. negative for all counts) does not indicate that the two counting efficiency curves might be equal. If the true counting efficiency is shifted within the shown error bars, for both curves the shift is in the same direction. The relative large error bars are mainly caused by subtracting the two roughly equal mass flow controller readings in order to derive the sampling volume flow (Sec. 4.3.2 and 5.3.3 Eq. 5.32).

Because the water peak (Fig. 5.5a) would be also counted by the CPC, only DMA selected particles were used for calibrating the OPC counting efficiency.

Consequently, the calibration could not be extended to particles larger than 500 nm (latex, 702 nm UTA). However, the OPC unit already reached the maximum counting efficiency of ~ 89% at a particle size of about 250 nm (UTA) and this asymptotic maximum counting efficiency stayed constant with increasing particle diameter.

Therefore, the counting efficiency was assumed to be at 89% for particles larger than 702 nm, too.

63  

Documento similar