• No se han encontrado resultados

Comparación de la improvisación de Hank Mobley y John Coltrane

CAPÍTULO 1: CONTEXTO HISTÓRICO

1.4 Comparación de la improvisación de Hank Mobley y John Coltrane

Chem. Listy 104, s634s639 (2010) ACP 2010  Súčasný stav a perspektívy analytickej chémie v praxi Posters %), and 3-NFt (99 %) were prepared by dissolving the

substances in 100 ml of methanol (gradient grade purity, Merck, Prague, CZ), the stock solution of 5-NQ (99 %) was prepared by dissolving the substance in 100 ml of deionized water. The concentration of all stock solutions was 1·10–3 mol l–1, the pure substances were supplied by

Sigma-Aldrich. All stock solutions were stored in glass vessels in the dark. It followed from a spectrophotometric study of the stability of all stock solutions that they are stable for at least 90 days.

Britton – Robinson (BR) buffers were prepared in a usual way, i.e. by mixing a solution of 0.04 mol l–1 in

phosphoric acid, 0.04 mol l–1 in acetic acid and 0.04 mol l–1

in boric acid with the appropriate amount of 0.2 mol l–1

sodium hydroxide solution. Borate buffer was prepared by mixing 0.05 mol l–1 of sodium tetraborate with appropriate

amount of 0.1 mol l–1 hydrochlorid acid to obtain pH 9.0.

Phosphate buffer was prepared by mixing 0.01 mol l–1

sodium dihydrogen phosphatewith appropriate amount of 0.2 mo L–1 sodium hydroxide solutionto obtain pH 7.0 (all

chemicals Lachema, Brno, CZ). Methanol (Merck, Prague, CZ) of gradient grade purity was used for mobile phase preparation. De-ionized water was produced by Milli-Qplus

system (Millipore, USA). Apparatus

Electrochemical detector used in FIA and HPLC measurements consisted of a platinum wire auxiliary elec- trode, silver/silver chloride (1 mol l–1 KCl) reference elec-

trode (both Monokrystaly, Turnov, CZ) and the working electrode, m AgSAE with the disc diameter 2.47 mm in three electrode arrangement used in “wall-jet” arrange- ment. m-AgSAE was prepared from a drawn-out glass tube, its narrow end was filled with a fine silver powder, amalgamated by liquid mercury and connected to an elec- tric contact16. The exact volume of mercury (0.4 μl if not

stated otherwise) to form the meniscus of m-AgSAE was applied by Hamilton syringe.

The FIA system consisted of linear high-pressure pump HPP 5001 (Laboratorní přístroje, Prague) and elec- trochemical and spectrometric detector. ED in “wall-jet” arrangement described earlier18 was used. Electrode sys-

tem was driven by potentiostat ADLC 2 (Laboratorní přístroje, Prague). The working m-AgSAE was kept at a constant potential so that the electrochemical cell served as an amperometric detector. The distance of electrode surface from the end of capillary was 0.5 mm. Samples were injected using six-way valve D (Ecom, Prague). Fol- lowing parameters for FIA determination were used: de- tection potential 1.6 V, flow rate 4 ml min–1 and injected

sample volume 0.1 ml.

Volume of spectrophotometric cell was 10 μl and optical path length was 5 mm. Output voltage corre- sponded to 5 V per 1 A.U. (1 mV = 2.10–4 A.U.). The sig-

nal of spectrophotometric detection was measured at con- stant wavelength 220 nm.

Program CSW32 version 1.4.11.06 (DataApex, Pra-

gue), driven under operational system Microsoft Windows 98 (Microsoft Corp.), was used for data collection.

The HPLC system consisted of high-pressure pump L-2130 HTA (Hitachi, Japan) governed by the software EZChrom Elite (Agilent Technologies, USA) working in Windows XP (Microsoft Corporation, USA). The column KROMASIL (250  4.6 mm, 7 μm) with reversed C18

phase (Phenomenex, Torrance, CA, USA) was used. Man- ual injection of samples degassed for 3 min by passing of nitrogen (purity 4.0, Linde, Prague, CZ) using 20 μl Rheo- dyne (IDEX Health & Science, Rohnert Park, CA, USA) injection valve was used. The diode array detector L-2450 (LaChrom Elite, Hitachi, USA) for measurement from 220 nm to 400 nm was used. The mobile phase was de- gassed by ultrasonic bath PS 02000A (Powersonic, USA) followed by passing nitrogen (purity 4.0, Linde, Prague, CZ) continuously for the whole period of measurement. The measurements were carried out at ambient laboratory temperature (22 °C).

The pH measurements were carried out by pH meter Jenway 4330 (Jenway, UK) with combined glass electrode (Ag/AgCl/3 mol l–1 KCl (type: 924005)). The pH meter

was calibrated with standard pH buffers (Sevac, Prague, CZ). pH values refer to those of the aqueous phase, pHf

values refer to those of the resulting pH of the mixtures of the aqueous phase with the organic solvent.

Procedures

In the case of flow injection analysis, 0.05 mol l–1

borate buffer, pH 9.0 was used as run buffer. More diluted analyte solutions were prepared by diluting of exact vol- ume of 5-NQ stock solution with run buffer. All samples were deaerated with N2 (5 min) before injection into flow

of run buffer. The m-AgSAE was activated before starting the work, as well as after every pause longer than one hour. The electrochemical activation of m-AgSAE was carried out by insert potential –2200 mV for 300 s in the stirred solution of 0.2 mol l–1 KCl and then washed by

distilled water. No other electrochemical pretreatment was performed prior each injection.

In the case of HPLC-ED, the mixture of 0.01 mol l–1

phosphate buffer, pH 7.0 : methanol (15:85, v/v) was used as mobile phase. More diluted analyte solutions were pre- pared by dilution of exact volume of the stock solutions with mobile phase. The flow rate Fm was set at 1 ml min–1

and the injected sample volume Vinj was 20 μl. The m-

AgSAE in HPLC experiments was activated by the same way as in FIA measurements. No other electrochemical pretreatment was performed prior each injection.

All calibration curves were measured in triplicate. The statistical parameters of calibration curves were calcu- lated using statistic software OriginPro 6.0 (OriginLab Corporation, USA). The significance of the intercepts of linear calibration dependences was tested by statistic soft- ware ADSTAT24, which was also used to calculate the

limit of quantitation in FIA measurements.

quantitation (LQ) were calculated from the peaks heights as

the concentration of an analyte which gave a signal three and ten times the background noise (S/N = 3; S/N = 10, respectively). All the statistical data are calculated for the level of significance α = 0.05.

Results and discussion

Flow injection analysis using m-AgSAE as amperometric sensor

First, the conditions for determination 5-NQ were optimized. 0.05 mol l–1 borate buffer pH 9.0 medium was

used as a run buffer for FIA determination of 5-NQ. The UV detector was also serially arranged prior to the electro- chemical detector for the sake of comparison. Optimal detection potential (Edet) inserted on m-AgSAE, flow rate

(Fm) of the run buffer and injection volume (Vinj) were

found during the process of optimization. Firstly, the influ- ence of the Edet on electrochemical signal was investigated.

Edet = –1.6 V was selected as optimum potential as it pro-

vides the highest signals. Further, the optimum flow rate was investigated (Fig. 1). Fm = 4 ml min–1 was set as the

maximum of peak heights achieved by electrochemical detector (and optimum for UV detector has the same value). The injected sample volume (Fig. 2) was set at 0.1 ml, which provides the highest signal, further not increas- ing when higher volumes are injected. This optimum value was reached for both detectors.

Calibration dependences were measured under opti- mized conditions in the range from 2·106 mol l1 to 1·104 mol l1 using ED and UV detector, they are linear within the whole concentration range. The peak heights were evaluated from the negative peak height obtained after injection of blank solution into FIA system. The record

peak of the ED is shown in Fig. 3. In Figure 3B detecting the lowest attainable concentration range from 2·106 mol l1 to 1·105 mol l1, there is a negative peak which is related to the presence of oxygen in the run buffer. The system is probably not totally hermetic and the freshly deaerated blank solution (and the samples) has lower content of oxy- gen than run buffer. This phenomenon was not observed for UV detector. The LQ achieved with UV (6·106 mol l1) and

ED (3·106 mol l1) are comparable.

The HPLC-ED method for the determination of 1-NP, 2-NF, 3-NFt, and 5-NQ

Firstly, the separation of 1-NP, 2-NF, 3-NFt, and 5- NQ was optimized using reversed C18 phase. Mobile phase

with high content of organic phase has to be used due to the extended aromatic system of studied NPAHs. Using 0.01 mol l–1 phosphate buffer, pH 7.0 : methanol (15:85, v/

v) mobile phase, baseline separation of tested analytes was achieved in fifteen minutes.

The detection potential Edet = –1.5 V was chosen

based on hydrodynamic voltammograms (Fig. 4) of tested analytes and the highest signal to noise ratio. The conse- quent decrease of peak currents Ip after the height maxi-

mum at Edet is reached is caused by an increase of the

background current due to the cathodic decomposition (i.e., hydrogen evolution) of the mobile phase. It proceeds at relatively negative potentials due to the high content of the organic phase in mobile phase. The noise of the system is independent of Edet and it is about 50 nA.

Calibration dependences were measured under opti- mized conditions within the range from 2·10–5 to 1·10–4

mol l–1 using electrochemical detector and within the range

from 3·10–7 to 1·10–4 mol l–1 using diode array detector

(DAD).

Parameters for the electrochemical detector are sum- marized in the Table I, limit of quantitation (10 S/N), is

0 4 8 -15000 -30000 -45000 IP , n A Fm, mL min-1 Fig. 1. Dependence of the peak heights Ip of 5-NQ (c = 1·10−4 mol l−1) on flow rate F

m (Edet = −1.6 V) for FIA-ED determina- tion of 5-NQ. Run buffer 0.05 mol l−1 borate buffer, pH 9.0;

injected volume Vinj = 100 l. Evaluated from peak heights

0 200 400 0 -20000 -40000 -60000 IP , n A Vinj, L

Fig. 2. Dependence of the peak heights Ip of 5-NQ (c = 1·10−4 mol l−1) on injected volume V

inj (Edet = −1.6 V) for FIA-ED determination of 5-NQ. Run buffer 0.05 mol l−1 borate buffer,

Chem. Listy 104, s634s639 (2010) ACP 2010  Súčasný stav a perspektívy analytickej chémie v praxi Posters

always 3.3 times higher than LD. The achieved LQs (from

3.0·10–5 mol l–1 for 5-NQ to 1.0·10–4 mol l–1 for 3-NFt) are

more than one order of magnitude than LQ reported above

for FIA-ED for 5-NQ (3·10–6 mol l–1). This may be caused

by higher noise using the high-pressure pump working in pulse regime in HPLC rather than the linear pump used for FIA.

Using DAD the optimum wavelength was found for each test substance. They are listed in the Table II together with the parameters of calibration dependences evaluated from both, peak areas and peak heights. The noise is inde- pendent on the wavelength and its value is about 8·10–5

A.U. The differences of LD calculated at optimum λ for

each compounds listed in Table II (from 0.13·10–6 mol l–1

for 5-NQ to 0.39·10–6 mol l–1 for 3-NFt) and for the com-

promise wavelenght of λ = 227 (from 0.32·10–6 mol l–1 for

5-NQ to 0.59·10–6 mol l–1 for 3-NFt) are about compara-

ble. The LDs were evaluated from peak heights.

0 3 6 9 0 -15000 -30000 0 3 6 9 -150 -300 t, min 1 2 3 4 5 A I, nA 2 1 I, nA t, min 3 4 5 6 B

Fig. 3. Peaks of 5-NQ recorded with FIA-ED with m-AgSAE detector in “wall-jet” arrangement for 5-NQ concentration: (A): 2

(1), 4 (2), 6 (3), 8 (4) and 10 (5) ·105 mol l–1, and (B): 0 (1), 2 (2), 4 (3), 6 (4), 8 (5) and 10 (6) ·106 mol l–1. Run electrolyte 0.05 mol l–1

borate buffer pH 9.0, Edet = –1.6 V, flow rate Fm = 4 ml min1, injected volume Vinj = 100 μl

-1,0 -1,2 -1,4 -1,6 -500 -1000 -1500 I p , nA Edet, V 1 2 3 4

Fig. 4. Dependence of peak heights Ip of 5-NQ (1), 2-NF (2), 1- NP (3), and 3-NFt (4) on detection potential Edet in HPLC-ED.

Concentration of each analyte 1·10–4 mol l–1. Mobile phase 0.01

mol l–1 phosphate buffer pH 7.0 : methanol (15:85; v/v),

Fm = 1 ml min–1, Vinj = 20 μl

Table I

Parameters of calibrations dependences for the determination of 5-NQ, 2-NF, 1-NP, and 3-NFt using HPLC-ED with “wall-jet” m-AgSAE detector. Mobile phase 0.01 mol l–1 phosphate buffer pH 7.0 : methanol (15:85; v/v), F

m = 1 ml min–1,

Vinj = 20 l, detection potential –1.5 V, evaluated from peak heights

Analyte Concentration range

[mol l–1] Slope[mA l mol–1] Correlationcoefficient [mol lLD –1] L[mol lQ –1]

5-NQ 20 – 100 –150 0.9917 9.2 30.4

2-NF 20 – 100 –99 0.9911 17.0 56.1

1-NP 20 – 100 –84 0.9952 28.0 92.4

Conclusion

In this study, we focused on the FIA-ED and HPLC- ED determination of selected nitrated aromatic compounds using an amperometric sensor in “wall-jet” detection cell.

FIA-ED was tested using 5-NQ. Using optimized conditions (run buffer 0.05 mol l1 borate buffer of pH 9.0, detection potential 1.6 V, flow rate 4 ml min1, injected

volume 0.1 ml), the LQ obtained for FIA-ED (3·106mol l1)

is comparable with UV detection (LQ = 6·106 mol l1).

Further, HPLC separation of 1-NP, 2-NF, 3-NFt, and 5-NQ using C18 phase was proposed in mobile phase con-

sisting of 0.01 mol l1 phosphate buffer pH 7.0: methanol 15:85 (v/v). Using electrochemical detection , LDs lying

mostly in the 105 mol l–1 concentration range were

achieved, which is about 2 orders of magnitude higher than using DAD when the optimum wavelengths for each analyte are used. In the following measurement we will focuse on the removal of the electronic noise from the measurement system, which contributes significantly to the relatively high LDs of electrochemical detection.

The project was supported by the Czech Ministry of Education, Youth and Sports (projects LC 06035, MSM 0021620857, and RP 14/63) and by Grant Agency of Charles University (project SVV 261 204).

REFERENCES

1. WHO: Selected Nitro- and Nitro-Oxy-Polycyclic Aro- matic Hydrocarbons. WHO, Geneva 2003.

2. Scheepers P. T. J., Bos R. P.: Int. Arch. Occup. Envi- ron. Health 64, 149 (1992).

3. Atkinson R., Arey J.: Environ. Health. Perspect. 102, 117 (1994).

4. Jacob J., Karcher W., Belliardo J. J., Dumler R., Boenke A.: Fresenius’ J. Anal. Chem. 340, 755 (1991).

5. Moreira J. C., Barek J.: Quim. Nova 18, 362 (1995). 6. Netto A. D. P., Moreira J. C., Dias A., Arbilla G.,

Ferreira L. F. V., Oliveira A. S., Barek J.: Quim. Nova 23, 765 (2000).

7. Zielinska B., Samy S.: Anal. Bioanal. Chem. 386, 883 (2006).

8. Kuo C. T., Chen H. W., Lin S. T.: Anal. Chim. Acta 482, 219 (2003).

9. Murayama M., Dasgupta K.: Anal. Chem. 68, 1226 (1996).

10. MacCrehan W. A., May W. E., D. Y. S., Benner S. D.: Anal. Chem. 60, 194 (1988).

11. Rosenkranz H. S.: Mutat. Res. 101, 1 (1982). 12. Galceran M. T., Moyano E.: Talanta 40, 615 (1993). 13. Jin Z., Rappaport S. M.: Anal. Chem. 55, 1778

(1983).

14. Novotny L., Yosypchuk B.: Chem. Listy 94, 1118 (2000).

15. Mikkelsen O., Schroder K.: Anal. Lett. 33, 3253 (2000).

16. Yosypchuk B., Novotny L.: Electroanalysis 14, 1733 (2002).

17. Yosypchuk B., Novotny L.: Crit. Rev. Anal. Chem. 32, 141 (2002).

18. Barek J., Fischer J., Navratil T., Peckova K., Yosyp- chuk B., Zima J.: Electroanalysis 19, 2003 (2007). 19. Yosypchuk B., Barek J.: Crit. Rev. Anal. Chem. 39,

189 (2009).

20. Peckova K., Vrzalova L., Bencko V., Barek J.: Collect. Czech. Chem. Commun. 74, 1697 (2009). 21. Danhel A., Shiu K. K., Yosypchuk B., Barek J.,

Peckova K., Vyskocil V.: Electroanalysis 21, 303 (2009).

Table II

Parameters of calibrations dependences for the determination of 5-NQ, 2-NF, 1-NP, and 3-NFt using HPLC-DAD. Mobile phase 0.01 mol l–1 phosphate buffer pH 7.0 : methanol (15:85; v/v), F

m = 1 ml min–1, Vinj = 20 l Analyte λ [nm] Concentration range [mol l–1] Slope a Correlation coefficient LD [mol l–1]

A - evaluated for peak height

5-NQ 220 0.3 – 100 1790 0.9998 0.13

2-NF 335 0.3 – 100 656 0.9997 0.36

1-NP 238 0.3 – 100 1080 0.9996 0.22

3-NFt 233 0.3 – 100 621 0.9993 0.39

B- evaluated for peak area

5-NQ 220 0.3 – 100 10510 0.9999 

2-NF 335 0.3 – 100 5620 0.9996 

1-NP 238 0.3 – 100 13220 0.9999 

Chem. Listy 104, s634s639 (2010) ACP 2010  Súčasný stav a perspektívy analytickej chémie v praxi Posters 22. Naughton A. B. J., Jespersen N. D.: Anal. Chim. Acta

251, 95 (1991).

23. Bleeker E. A. J., Wiegman S., de Voogt P., Kraak M., Leslie H. A., de Haas E., Admiraal W. S.: Reviews of Environmental Contamination and Toxicology. Springer, New York 2002.

24. Meloun M., Militký J., Forina M: Chemometrics for Analytical Chemistry. Ellis Horwood, Chichester 1992.

Documento similar