A.41.3 Ophthalmic Photography 8 Bit Image IOD Modules

Table A.41-1 OPHTHALMIC PHOTOGRAPHY 8 BIT IMAGE IOD MODULES

IE Module Reference Usage

Patient Patient C.7.1.1 ALWAYS

Study General Study C.7.2.1 ALWAYS

Series General Series C.7.3.1 ALWAYS

Ophthalmic Photography Series C.8.17.1 ALWAYS Frame of

Reference

Synchronization C.7.4.2 ALWAYS

Equipment General Equipment C.7.5.1 ALWAYS

Image General Image C.7.6.1 ALWAYS

Image Pixel C.7.6.3 ALWAYS

Acquisition Context C.7.6.14 ALWAYS

Enhanced Contrast/Bolus C 7.6.4b CONDITIONAL - Included if contrast was administered

Cine C.7.6.5 ALWAYS

Multi-frame C.7.6.6 ALWAYS

Ophthalmic Photography Image C.8.17.2 ALWAYS

Ocular Region Imaged C.8.17.5 ALWAYS

Ophthalmic Photography Acquisition Parameters

C.8.17.4 ALWAYS Ophthalmic Photographic

Parameters

C.8.17.3 ALWAYS

SOP Common C.12.1 ALWAYS

Tag VR Name Value PoV Source

Information Entity 'Patient' Module 'Patient'

(0010,0010) PN Patient's Name Patient's full name. VNAP MWL, USER

(0010,0020) LO Patient ID Primary hospital

identification number or code for the patient.

VNAP MWL, USER

(0010,0021) LO Issuer of Patient ID

Identifier of the Assigning Authority that issued the

Patient ID. ANAP MWL

(0010,0030) DA Patient's Birth Date Birth date of the patient. VNAP MWL, USER

(0010,0040) CS Patient's Sex

Sex of the named patient.

Enumerated Values: M =

male F = female O = other VNAP MWL, USER

(0010,1000) LO Other Patient IDs Other identification numbers or codes used to identify the patient.

ANAP MWL

(0010,4000) LT Patient Comments User-defined additional information about the patient.

ANAP MWL, USER

Document: DICOM_Conformance_Statement_VISUCAM_224_524_6.0.doc Page 23 of 60 Copyright: © Carl Zeiss Meditec AG EN_31_200_0031I Revision: 1.0 Information Entity 'Study'

Module 'General Study'

(0008,0020) DA Study Date Date the Study started. ALWAYS AUTO (0008,0030) TM Study Time Time the Study started. ALWAYS AUTO

(0008,0050) SH Accession Number

A RIS generated number that identifies the order for

the Study. VNAP MWL

(0008,0090) PN Referring Physician's Name Name of the patient's

referring physician VNAP MWL

(0008,1030) LO Study Description

Institution-generated description or classification of the Study (component) performed.

In the scheduled case the value is copied from Modality Worklist attribute Requested Procedure Description.

ANAP MWL

(0008,1032) SQ Procedure Code Sequence

A Sequence that conveys the type of procedure performed. One or more Items may be included in this Sequence.

Included macro 'Code Sequence Macro', context 'No Baseline Context ID is defined.'

ANAP MWL

>(0008,0100) SH Code Value See Section 8.1. Required if

a sequence item is present. ALWAYS MWL >(0008,0102) SH Coding Scheme

Designator See Section 8.2. Required if

a sequence item is present. ALWAYS MWL

>(0008,0103) SH Coding Scheme Version

See Section 8.2. Required if a sequence item is present and the value of Coding Scheme Designator

(0008,0102) is not sufficient to identify the Code Value (0008,0100)

unambiguously.

ANAP MWL

>(0008,0104) LO Code Meaning See Section 8.3. Required if

a sequence item is present. ALWAYS MWL

(0008,1048) PN Physician(s) of Record

Names of the physician(s) who are responsible for overall patient care at time of Study (see Section C.7.3.1 for Performing Physician)

Requesting physician’s name is copied from MWL.

ANAP MWL, AUTO

(0008,1110) SQ Referenced Study Sequence

A sequence that provides reference to a Study SOP Class/Instance pair. The sequence may have zero or

ANAP MWL

Document: DICOM_Conformance_Statement_VISUCAM_224_524_6.0.doc Page 24 of 60 Copyright: © Carl Zeiss Meditec AG EN_31_200_0031I Revision: 1.0

more Items.

>(0008,1150) UI Referenced SOP Class UID

Uniquely identifies the referenced SOP Class.

Required if Referenced Study Sequence (0008,1110) is sent.

ANAP MWL

>(0008,1155) UI Referenced SOP Instance UID

Uniquely identifies the referenced SOP Instance.

Required if Referenced Study Sequence (0008,1110) is sent.

ANAP MWL

(0020,000D) UI Study Instance UID

Unique identifier for the Study.

In the unscheduled case VISUCAM uses a constant prefix of

„1.2.276.0.75.2.1.20.0.1.“

followed by a date/time stamp and a machine specific identifier.

ALWAYS AUTO, MWL

(0020,0010) SH Study ID User or equipment

generated Study identifier. ALWAYS AUTO Module 'Patient Study'

Information Entity 'Series' Module 'General Series'

(0008,0021) DA Series Date Date the Series started. ALWAYS AUTO (0008,0031) TM Series Time Time the Series started. ALWAYS AUTO (0008,1050) PN Performing

Physicians' Name

Name of the physician(s)

administering the Series. ANAP CONFIG (0008,1070) PN Operators' Name Name(s) of the operator(s)

supporting the Series. ANAP CONFIG

(0018,0015) CS Body Part Examined

“HEAD”

ALWAYS AUTO

(0018,1030) LO Protocol Name

User-defined description of the conditions under which the Series was performed.

Note: This attribute conveys series-specific protocol identification and may or may not be identical to the one presented in the Performed Protocol Code Sequence (0040,0260).

EMPTY

(0018,5100) CS Patient Position

Patient position descriptor relative to the equipment.

Required for CT and MR images; shall not be present if Patient Orientation Code Sequence (0054,0410) is present; may be present otherwise. See C.7.3.1.1.2 for Defined Terms and

EMPTY

Document: DICOM_Conformance_Statement_VISUCAM_224_524_6.0.doc Page 25 of 60 Copyright: © Carl Zeiss Meditec AG EN_31_200_0031I Revision: 1.0

further explanation.

(0020,000E) UI Series Instance UID

VISUCAM uses a constant prefix of

„1.2.276.0.75.2.1.20.0.2.“

followed by a date/time stamp and a machine specific identifier.

Performed Procedure Step

started. ALWAYS AUTO

(0040,0245) TM Performed Procedure Step Start Time

Time on which the

Performed Procedure Step started.

ALWAYS AUTO

(0040,0253) SH Performed Procedure Step ID

User or equipment

generated identifier of that part of a Procedure that has been carried out within this step. description or classification of the Procedure Step that was performed.

EMTPY

(0040,0275) SQ Request Attributes Sequence

Sequence that contains attributes from the Imaging Service Request. The sequence may have one or more Items.

Included macro 'Request Attributes Macro', context 'No Baseline Context IDs defined' administrative description or classification of Requested Procedure. description or classification of the Scheduled Procedure Step to be performed.

ANAP MWL

>(0040,0008) SQ Scheduled Protocol Code Sequence

Sequence describing the Scheduled Protocol following a specific coding scheme.

This sequence contains one or more Items.

Included macro 'Code Sequence Macro', context 'Context ID may be defined in the macro invocation.'

ANAP MWL

>>(0008,0100) SH Code Value See Section 8.1. Required if

a sequence item is present. ALWAYS MWL >>(0008,0102) SH Coding Scheme

Designator

See Section 8.2. Required if

a sequence item is present. ALWAYS MWL >>(0008,0103) SH Coding Scheme See Section 8.2. Required if ANAP MWL

Document: DICOM_Conformance_Statement_VISUCAM_224_524_6.0.doc Page 26 of 60 Copyright: © Carl Zeiss Meditec AG EN_31_200_0031I Revision: 1.0

Version a sequence item is present and the value of Coding Scheme Designator

(0008,0102) is not sufficient to identify the Code Value (0008,0100)

unambiguously.

>>(0008,0104) LO Code Meaning See Section 8.3. Required if

a sequence item is present. ALWAYS MWL >(0040,0009) SH Scheduled

Procedure Step ID Identifier that identifies the

Scheduled Procedure Step. ANAP MWL

>(0040,1001) SH Requested Procedure ID

Identifier that identifies the Requested Procedure in the Imaging Service Request.

ANAP MWL

Module 'Ophthalmic Photography Series'

(0008,0060) CS Modality “OP” ALWAYS AUTO

Information Entity 'Frame of Reference' Module 'Synchronization'

(0018,106A) CS Synchronization Trigger

“NO TRIGGER” - data acquisition is not

synchronized by common channel or trigger.

ALWAYS AUTO

(0018,1800) CS Acquisition Time Synchronized

“N “ - Acquisition Datetime (0008,002A) is not

synchronized with external time reference.

ALWAYS AUTO

(0020,0200) UI Synchronization Frame of Reference UID

The UID is valid from Instrument Application’s start-up til shut down. The internal clock can be adjusted by service personel only.

After adjusting the internal clock, the service member restarts the Instrument Application. So it is guaranteed the UID represents one single Synchronization Frame of Reference.

VISUCAM uses a constant prefix of

“1.2.276.0.75.2.5.10.1.2.”

followed by a date/time stamp and a machine specific identifier.

ALWAYS AUTO

Information Entity 'Equipment' Module 'General Equipment '

(0008,0070) LO Manufacturer “Carl Zeiss Meditec AG” ALWAYS AUTO

(0008,0080) LO Institution Name Institution where the

equipment that produced the composite instances is located.

ANAP CONFIG

(0008,0081) ST Institution Address

Mailing address of the institution where the

equipment that produced the ANAP CONFIG

Document: DICOM_Conformance_Statement_VISUCAM_224_524_6.0.doc Page 27 of 60 Copyright: © Carl Zeiss Meditec AG EN_31_200_0031I Revision: 1.0

composite instances is located.

(0008,1010) SH Station Name

User defined name identifying the machine that produced the

composite instances. ANAP CONFIG

(0008,1040) LO Institutional Department Name

Department in the institution where the equipment that produced the composite instances is located.

ANAP CONFIG

(0008,1090) LO Manufacturer's Model Name

Manufacturer's model name of the equipment that produced the composite instances.

“VISUCAM 224” or “VISUCAM 524”

ALWAYS AUTO

(0018,1000) LO Device Serial Number

Manufacturer's serial number of the equipment that

produced the composite instances. Note: This identifier corresponds to the device that actually created the images, such as a CR plate reader or a CT console, and may not be sufficient to identify all of the equipment in the imaging chain, such as the generator or gantry or plate.

ALWAYS AUTO

(0018,1020) LO Software Versions

Manufacturer's designation of software version of the equipment that produced the composite instances.

Always “6.0.0.39866” and higher versions "6.0.x.y"

where x denotes a patch version and y denotes a build version

ALWAYS AUTO

(0018,1050) DS Spatial Resolution

The inherent limiting resolution in mm of the acquisition equipment for high contrast objects for the data gathering and reconstruction technique chosen. If variable across the images of the series, the value at the image center.

VNAP AUTO

Information Entity 'Image' Module 'General Image'

(0008,0022) DA Acquisition Date

The date the acquisition of data that resulted in this

image started ALWAYS AUTO

(0008,0032) TM Acquisition Time

The time the acquisition of data that resulted in this

image started ALWAYS AUTO

(0020,0020) CS Patient Orientation

Patient direction of the rows and columns of the image.

Required if image does not require Image Orientation

ALWAYS AUTO

Document: DICOM_Conformance_Statement_VISUCAM_224_524_6.0.doc Page 28 of 60 Copyright: © Carl Zeiss Meditec AG EN_31_200_0031I Revision: 1.0

(Patient) (0020,0037) and Image Position (Patient) (0020,0032). See C.7.6.1.1.1 for further explanation. Note: IOD's may have attributes other than Patient Orientation, Image Orientation, or Image Position (Patient) to describe orientation in which case this attribute will be zero length.

Always “L/F”

(0020,4000) LT Image Comments User-defined comments

about the image VNAP USER

Module 'Acquisition Context'

(0040,0555) SQ Acquisition Context Sequence

A sequence of Items that describes the conditions present during the

acquisition of the data of the SOP Instance. Zero or more items may be included in this sequence.

ALWAYS AUTO

>(0040,08EA) SQ Measurement Units Code Sequence

Units of measurement. Only a single Item shall be permitted in this Sequence.

Required if Numeric Value (0040,A30A) is sent. Shall not be present otherwise.

Included macro 'Code Sequence Macro', context 'Baseline Context ID is 82. '

ANAP AUTO

>>(0008,0100) SH Code Value See Section 8.3. Required if

a sequence item is present. ALWAYS AUTO >>(0008,0102) SH Coding Scheme Designator See Section 8.3. Required if

a sequence item is present. ALWAYS AUTO

>>(0008,0103) SH Coding Scheme Version

See Section 8.3. Required if a sequence item is present and the value of Coding Scheme Designator

(0008,0102) is not sufficient to identify the Code Value

ANAP AUTO

Document: DICOM_Conformance_Statement_VISUCAM_224_524_6.0.doc Page 29 of 60 Copyright: © Carl Zeiss Meditec AG EN_31_200_0031I Revision: 1.0

(0008,0100) unambiguously.

>>(0008,0104) LO Code Meaning See Section 8.3. Required if

a sequence item is present. ALWAYS AUTO

>(0040,A043) SQ Concept Name Code Sequence

A concept that constrains the meaning of (i.e. defines the role of) the Observation Value. The "Name"

component of a Name/Value pair. This sequence shall contain exactly one item.

Included macro 'Code Sequence Macro', context 'No Baseline Context is defined.'

ANAP AUTO

>>(0008,0100) SH Code Value See Section 8.3. Required if

a sequence item is present. ALWAYS AUTO >>(0008,0102) SH Coding Scheme

Designator

See Section 8.3. Required if

a sequence item is present. ALWAYS AUTO

>>(0008,0103) SH Coding Scheme Version

See Section 8.3. Required if a sequence item is present and the value of Coding Scheme Designator

(0008,0102) is not sufficient to identify the Code Value (0008,0100)

unambiguously.

ANAP AUTO

>>(0008,0104) LO Code Meaning See Section 8.3. Required if

a sequence item is present. ALWAYS AUTO

>(0040,A30A) DS Numeric Value

This is the Value component of a Name/Value pair when the Concept implied by Concept Name Code Sequence (0040,A043) is a set of one or more numeric values. Required if the value that Concept Name Code Sequence (0040,A043) requires (implies) is a set of one or more integers or real numbers. Shall not be present otherwise.

ANAP AUTO

>(0040,A122) TM Time

This is the Value component of a Name/Value pair when the Concept implied by Concept Name Code Sequence (0040,A043) is a time.

Note

The purpose or role of the time value could be specified in Concept Name Code Sequence (0040,A043).

Required if the value that Concept Name Code

ANAP AUTO

Document: DICOM_Conformance_Statement_VISUCAM_224_524_6.0.doc Page 30 of 60 Copyright: © Carl Zeiss Meditec AG EN_31_200_0031I Revision: 1.0

Sequence (0040,A043) requires (implies) is a time.

Shall not be present otherwise.

Attribute exists for the concepts “FA start time” and

“ICG start time”

>(0040,A168) SQ Concept Code Sequence

This is the Value component of a Name/Value pair when the Concept implied by Concept Name Code Sequence (0040,A043) is a Coded Value.

Only a single Item shall be included in this sequence.

Required if Date (0040,A121), Time

(0040,A122), Person Name (0040,A123), Text Value (0040,A160), and the pair of Numeric Value (0040,A30A) and Measurement Units Code Sequence (0040,08EA) are not present.

Attribute exits for the concepts “Imaging agent”,

“Sensor” and “Protocol name”

ANAP AUTO

>>(0008,0100) SH Code Value See Section 8.3. Required if

a sequence item is present. ALWAYS AUTO >>(0008,0102) SH Coding Scheme Designator See Section 8.3. Required if

a sequence item is present. ALWAYS AUTO

>>(0008,0103) SH Coding Scheme Version

See Section 8.3. Required if a sequence item is present and the value of Coding Scheme Designator

(0008,0102) is not sufficient to identify the Code Value (0008,0100)

unambiguously.

ANAP AUTO

>>(0008,0104) LO Code Meaning See Section 8.3. Required if

a sequence item is present. ALWAYS AUTO Module 'Enhanced Contrast/Bolus'

(0018,0012) SQ Contrast/Bolus Agent Sequence

Contains zero or one item.

Sequence that identifies one or more contrast agents administered prior to or during the acquisition. Shall contain one or more Items.

Included macro 'Code Sequence Macro', context 'Baseline Context ID is 12.' Only included in case of FA

ANAP AUTO

Document: DICOM_Conformance_Statement_VISUCAM_224_524_6.0.doc Page 31 of 60 Copyright: © Carl Zeiss Meditec AG EN_31_200_0031I Revision: 1.0

or ICG acquisition.

>(0008,0100) SH Code Value “C-B02CC” for FA or

“C-B0156” for ICG ALWAYS AUTO >(0008,0102) SH Coding Scheme Designator “SRT” ALWAYS AUTO

>(0008,0103) SH Coding Scheme

Version “20040921” ALWAYS AUTO

>(0008,0104) LO Code Meaning

“Fluorescein” for FA or

“Indocyanin green” for ICG ALWAYS AUTO

>(0018,0014) SQ Contrast/Bolus Administration Route Sequence

Contains one item.

Sequence that identifies the route of administration of contrast agent. Shall contain exactly one Item.

Included macro 'Code Sequence Macro', context 'Baseline Context ID is 11.' Only included in case of FA or ICG acquisition.

ALWAYS AUTO

>>(0008,0100) SH Code Value “G-D101” ALWAYS AUTO >>(0008,0102) SH Coding Scheme

Designator “SNM3” ALWAYS AUTO

>>(0008,0104) LO Code Meaning “Intravenous route” ALWAYS AUTO >(0018,1041) DS Contrast/Bolus

Volume

Exists for conformance, but

empty. EMPTY AUTO

>(0018,1041) DS Contrast/Bolus Volume Exists for conformance, but

empty. EMPTY AUTO

>(0018,9337) US Contrast/Bolus Agent Number

“1” because at maximum there is one item in this

Exists for conformance, but empty.

Active ingredient of agent.

Zero or more Items may be included in the Sequence.

Included macro 'Code Sequence Macro', context 'Baseline Context ID is 13.'

EMPTY AUTO

>(0018,9340) SQ Contrast Administration Profile Sequence

Contains one item.

Sequence that describes one or more phases of contrast administered. If present, shall contain one or more Items.

ALWAYS AUTO

>>(0018,1041) DS Contrast/Bolus Volume

Exists for conformance, but empty.

Volume administered during

EMPTY AUTO

Document: DICOM_Conformance_Statement_VISUCAM_224_524_6.0.doc Page 32 of 60 Copyright: © Carl Zeiss Meditec AG EN_31_200_0031I Revision: 1.0

this phase in milliliters of diluted contrast agent.

>>(0018,1042) TM Contrast/Bolus Start Time

Time of start of

administration. ALWAYS AUTO Module 'Cine'

(0018,1063) DS Frame Time

Nominal time (in msec) per individual frame. See C.7.6.5.1.1 for further explanation. Required if Frame Increment Pointer (0028,0009) points to Frame Time.

“0”

ALWAYS AUTO

Module 'Multi-frame'

(0028,0008) IS Number of Frames “1” – the VISUCAM does not

create multi-frame images ALWAYS AUTO (0028,0009) AT Frame Increment

Pointer “(0018,1063)” ALWAYS AUTO

Module 'Ophthalmic Photography Image'

(0008,0008) CS Image Type

Image identification characteristics.

Multi-value attribute containing 4 values:

1) Pixel Data Characteristics

• “ORIGINAL” for original acquired images

• “DERIVED” for any derived image

2) Patient Examination Characteristics

• “PRIMARY” or

“SECONDARY”

3) Modality Specific Characteristics

• “MONTAGE” for panorama images

• empty otherwise 4) Implementation specific identifiers(according to selected capture mode)

• “COLOR”, “REDFREE”,

“RED”, “BLUE”, “AF”, “FA”1,

“ICG”1

1: only available for VISUCAM 524

ALWAYS AUTO

(0008,0023) DA Content Date The date the image pixel

data creation started. ALWAYS AUTO

(0008,002A) DT Acquisition Datetime The date and time that the acquisition of data started.

Note: The synchronization of

ALWAYS AUTO

Document: DICOM_Conformance_Statement_VISUCAM_224_524_6.0.doc Page 33 of 60 Copyright: © Carl Zeiss Meditec AG EN_31_200_0031I Revision: 1.0

this time with an external clock is specified in the synchronization Module in Acquisition Time

Synchronized (0018,1800).

Required if Image Type (0008,0008) Value 1 is ORIGINAL. May be present otherwise.

(0008,0033) TM Content Time The time the image pixel

data creation started. ALWAYS AUTO

(0008,2112) SQ Source Image Sequence

A Sequence that identifies the set of Image SOP Class/Instance pairs of the Images that were used to derive this Image. Required if Image Type Value 1 is DERIVED. Zero or more items may be present in the sequence. See C.7.6.1.1.4 for further explanation.

Included macro 'Image SOP Instance Reference Macro', context ''

Always empty sequence.

EMPTY

“MONOCHROME2” – for monochrome images.

“RGB” – for color images.

“YBR_FULL_422” – for JPEG Baseline compressed images.

ALWAYS CONFIGAUTO

(0028,0006) US Planar Configuration

“0” - color-by-pixel Exists if Samples per Pixel (0028,0002) has a value greater than 1.

ANAP AUTO

(0028,0030) DS Pixel Spacing

Present because Acquisition Device Type Code Sequence (0022,0015) contains an item with the value (SRT, R-1021A, "Fundus Camera").

Values depend on choosen visible angle.

Specifies whether an Image has undergone lossy compression. “00” - Image has not been subjected to lossy compression. “01” -

ALWAYS CONFIG AUTO

Document: DICOM_Conformance_Statement_VISUCAM_224_524_6.0.doc Page 34 of 60 Copyright: © Carl Zeiss Meditec AG EN_31_200_0031I Revision: 1.0

Image has been subjected to lossy compression.

(0028,2112) DS Lossy Image Compression Ratio

Describes the approximate lossy compression ratio(s) that have been applied to this image. See C.7.6.1.1.5 for further explanation. May be multivalued if successive lossy compression steps have been applied. Notes: 1.

For example, a compression ratio of 30:1 would be described in this Attribute with a single value of 30. 2.

For historical reasons, the lossy compression ratio may also be described in

Derivation Description (0008,2111). Required if Lossy Image Compression (0028,2110) has a value of

"01".

“IDENTITY” – Exists if Photometric Interpretation (0028,0004) is

“MONOCHROME2”

ANAP AUTO

Module 'Ocular Region Imaged'

(0008,2218) SQ Anatomic Region Sequence

One item present

Sequence that identifies the anatomic region of interest in this Instance (i.e. external anatomy, surface anatomy, or general region of the body). Only a single Item shall be permitted in this sequence.

Laterality of object imaged (as described in Anatomic Region Sequence

(0008,2218)) examined.

Enumerated Values: R = right eye L = left eye B = both left and right eye Shall be consistent with any laterality information contained in Primary Anatomic Structure Modifier

ALWAYS AUTO

Document: DICOM_Conformance_Statement_VISUCAM_224_524_6.0.doc Page 35 of 60 Copyright: © Carl Zeiss Meditec AG EN_31_200_0031I Revision: 1.0

Sequence (0008,2230), if present. Note: Laterality (0020,0060) is a Series level Attribute and must be the same for all Images in the Series. Since most

Ophthalmic Photographic Image studies contain images of both eyes, the series level attribute will rarely be present.

Module 'Ophthalmic Photography Acquisition Parameters'

(0022,0005) CS

Patient Eye Movement Commanded

Empty, exists because of

conformance. EMPTY

(0022,000A) FL Emmetropic Magnification

Empty, exists because of conformance.

Emmetropic Magnification

Emmetropic Magnification

Documento similar