Capítulo VIII. Resumen de la Inversión
Anexo 21. Matriz de Riesgos
From Experiment 2, fluorescent images of platelet adhesion each coating after blood exposure are
in Fig. A. 1.15, taken at 40X magnification to visualize a wide view of platelet distribution on the
surfaces. On the unmodified PDMS surface (Fig. A. 1.15a), platelets can be seen distributed across the
surface at a consistent, low density. Some platelets appear stained with only FITC, indicating adhesion
but no activation, while some are also cross-stained with Cy-3, pointing to activation. Of the coatings, HA
(Fig. A. 1.15b) shows the least platelet adhesion/activation, with only a few platelets present on the
146 activation compared to the unmodified PDMS surface, and the PAA-His surface (Fig. A. 1.15d) shows a
modest increase. In particular, the Hemin surface is covered by a high density of cross-stained platelets,
some of which have started to form larger aggregates. IHC imaging was not conducted on samples from
Experiment 1.
Fig. A. 1.15: Experiment 2 – 40X magnification IHC cross-stain of platelet adhesion (CD41, FITC green)
and platelet activation (CD62p, Cy-3 red) on membranes removed from the in vivo devices after nearly 4 hours of blood contact. Coatings are: (a) unmodified PDMS, (b) HA, (c) Hemin, and (d) PAA-His.
DISCUSSION
Considering the data sets from all experiments, the biomolecule coatings strongly altered the
hemocompatibility of the Si-PDMS membrane surfaces. Whether the benefit was positive or negative, the
coatings generally showed consistent trends between in vitro wettability and in vivo hemocompatibility,
and mostly differed from the unmodified PDMS surfaces. Between the two experiments, it is unclear if
the coatings have a detrimental impact on the inherent gas permeability of the Si-PDMS membranes.
147 evidence. Although the coatings in Experiment 2 consistently decreased the in vitro O2 flux through the
membranes, it is unclear if this trend was also observed in vivo due to the high measurement error. As
such, the biomolecule coatings demonstrated a clear hemocompatibility benefit compared to unmodified
PDMS, but were unable to show any blood-gas exchange improvements on gas exchange membranes.
The presence of the biomolecule coatings on the surface was confirmed by contact angle changes.
Characterization of the coatings was largely consistent with prior research on these coatings. To validate
the presence of the coatings on the surface, two additional characterization techinques were employed in
this study: colorimetric staining for amine and carboxyl groups and x-ray photoelectron spectroscopy.
However, these datasets were not sufficiently complete to include in this analysis. As such, coating
characterization in this study is currently limited to contact angle analysis. Still, other characterization
techniques may have been useful to further define the functionality of the coatings. Data on thickness and
conformity via ellipsometry could have identified possible links between coating thickness and
thrombogenicity. Atomic force microscopy could have been used to image the topography of the coated
surfaces, since surface roughness is also linked to thrombogenicity. Additionally, a technique to quantify
the surface charge of the biomolecule coatings could have verified the charge state of the coatings under
physiologic conditions.
Comparison of the characterization and hemocompatibility data revealed some trends consistent
with all data sets. Namely, the HA coating was the most hydrophilic, and also the most resistant to cell
and platelet adhesion. This finding was consistent between Experiments 1 and 2. Conversely, the most
hydrophobic coatings – PAA-Lys and Hemin – were more thrombogenic than HA, but to different
extents. PAA-His was intermediate in both regards. These comparisons indicated a strong connection
between hydrophilicity and hemocompatibility with the coatings. However, if hydrophobicity alone
determined hemocompatibility, the unmodified PDMS and PAA-Lys surfaces should have been as
148 unmodified PDMS material control despite being equally hydrophobic. Also, in Experiment 2, the
unmodified PDMS surface attracted relatively few cells and platelets, comparable to the PAA-His surface
and far less than Hemin. As such, factors beyond wettability likely contributed towards
hemocompatibility, although hydrophilicity was a clear factor in preventing thrombosis.
Between all data sets, the influence of the coatings on gas permeability was unclear. The two in
vitro data sets indicated opposite conclusions. Whereas Experiment 1 showed no significant difference
with coating application, Experiment 2 showed a clear reduction in permeability with the application of
any coating to the surface. Normalizing the flux measured through unmodified SµM-PDMS membranes
from Chapter 2 (Fig. 2.7) to surface area, the O2 flux measured in prior studies at 10 mL/min flow rate
was 306±35.2 mL/min/m2. Both datasets from Experiment 1 and 2 (Fig. A. 1.6 and Fig. A. 1.7) showed
far less flux on the unmodified membranes than this study at the same flow rate, although the membranes
used in these studies had different pore designs that may have influenced the normalization. Still, the flux
measured in Experiment 2 was much closer to this previous measurement than Experiment 1, reaching
close to the same flux at a higher water flow rate. Given the increased number of data points in
Experiment 2, as well as the use of higher flow rates, the in vitro dataset in Experiment 2 is likely more
reflective of the coatings’ effect on the inherent gas permeability of the membranes. As such, the coatings appear to strongly decrease gas permeability of SµM-PDMS membranes.
The reduction seen with DETA application alone may be the most significant contributor to the
overall lower permeability of the coatings, since most of the coatings did not decrease flux further than
DETA alone. One possibility for this observation may be related to the effects of the plasma treatment
required to apply the DETA coating. Plasma treatment of PDMS causes the formation of a stiff, glassy
SiO2 layer on the PDMS surfaces, which has been shown to strongly reduce the gas permeability of
PDMS 115,117. Usually, this SiO
2 layer dissipates over time due to surface rearrangement of the PDMS
149 microns thick, unlike the ultrathin spin-coated layer on the SµM-PDMS. It is possible that an SiO2 layer
on SµM-PDMS membranes may not experience much surface rearrangement over time due to its
thinness, and lack of sufficient PDMS required for rearrangement. Therefore, it is possible that the gas
permeability reduction associated with plasma treatment may linger on SµM-PDMS membranes for an
indefinite amount of time. Plasma-treated PDMS without DETA application was not independently
evaluated in this experiment, so further studies would need to confirm this observation.
The in vivo datasets from Experiments 1 and 2 were also unclear as to the impact of the coatings
on gas exchange. Neither experiment was able to demonstrate oxygenation, unlike in Chapter 3. Lack of
demonstrable oxygenation in both experiments is likely due to the instability of the ventilation regimen of
the pigs. Unstable control over the incoming venous blood, and a limited collection of data points per
device, resulted in high variance in O2 flux measurements. Similarly, CO2 flux measurements were
affected by this instability in ventilation, particularly in Experiment 2 with the wide variation in blood pH
(7.3-7.7 over the course of the experiment). Still, the extremely high CO2 removal seen in Experiment 1
was striking, especially compared to past results from Chapter 5 under similar hypercarbic conditions.
This CO2 removal vastly exceeds data from Arazawa, et al. (411 mL/min/m2), although – given the lack
of a control uncoated device in this experiment – it is unclear if the coatings were the main driver of the
high CO2 removal. When a control device was added in Experiment 2, no significant difference in CO2
removal was seen between uncoated and coated devices, possibly attributable to the inconsistency in the
pig’s ventilation. Considering all of the in vivo datasets, the impact of the coatings on in vivo gas exchange cannot be determined without additional experiments.
Future in vivo experiments would need to significantly modify the animal handling procedures in
order to maintain consistent O2 and CO2 control over the incoming venous blood from the pig. In
Experiments 1 and 2, the pig was ventilated with only O2 because equipment to mix multiple gases for the
150 venous blood (SO2 ~100%), and the subsequent hand bag ventilation regimen to create a slightly hypoxic
state. Titration of the inhaled ventilator gas with room air or N2 would prevent this venous hyperoxia, and
allow for controllable hypoxia within the venous blood. If such a ventilation regimen is used, better
quality in vivo data could be collected to demonstrate O2 and CO2 transport.
CONCLUSION
Ultimately, the biomolecule coatings on SµM-PDMS membranes did significantly improve the
hemocompatibility of the surfaces, but were unable to demonstrate consistent impacts on gas exchange
through the membranes. Of the coatings, the HA coating proved to be the most hydrophilic and
hemocompatible when exposed to flowing blood. Oxygen flux in vitro may have been significantly
lessened by the application of the coatings, possibly due to the plasma treatment required for surface
modification. However, in vivo no difference was observed in oxygenation with or without coatings.
Strikingly high CO2 removal was seen in one of the experiments, signaling significant promise for the
coatings in a CO2 removal device. However, this result was not validated by the second experiment. Both
of the experiments were impacted by the instability of the in vivo setup, and so neither experiment can
conclusively demonstrate the impact of the coatings on in vivo gas exchange. Nonetheless, this work
raises significant interest in the use of these coatings for improving PDMS-based oxygenator
hemocompatibility, and possibly the gas exchange efficiency through the membranes. Future work will
focus on increasing sample numbers to improve overall data quality for characterization and membrane
151