across opposite the sample for light access .
4 . 5 Polarised Light Observation
Study using polarised light is a very powerful tool with crystals of non-cubic symmetry . The purity of polarisation depends on both the quality and the orientation of the polar iser , so it is pointless to have one of these quantities
very good if the other does not match it .
The worst constraint on the addition of a polariser to this system was the limited space of the Bausch and Lomb
sample compartment . This and the convergence angle of the beam precluded the use of a polarising prism . It was decided therefore to use as analyser a piece of sheet polaroid mounted on the tip of the dewar. This had the advantage that align ment could be done independently of the rest of the optical system , and was correct for both spectrographs .
Two polarisation directions at right angles were required , both prealigned . This could be achieved most simply in a
minimum of space by mounting on one of the dewar windows a slide holding two aligned polaroids , as in Figure 4 . 4 .
The polaroid used was a type which in the extinction position gave a dark blue colour. To test its behaviour quantitatively , its spectrum was run in the Bausch and Lomb . It was known that the beam was roughly polarised horizontally ,
so to get both spectra a 45° setting was used . Crossed in
the range 5500 - 6800 A . This covers the region of interest nicely . A rise centred on 4300 A explains the blue colour . Figure 4 . 5 shows this, and also the absorb ance of a single polaroid in the two positions .
Since mounting small crystals was difficult enough
without lining the optic axis to an exact angle, the crystal was mounted to about the correct angle, and the polaroids
aligned to it . Each was crossed with a third piece on the other side of the dewar by observing extinction of light passing round the sample holder, and these were rotated
as a pair till extinction with the crystal was also observed . Late in the experimental work a red filter was used to give a better extinction . If the angle of misalignment is M,
the intensity ratio of wanted to unwanted components is tan2M. M was better than 1°, giving ratio 0 . 0003 . A small amount of wobble in the slide, say 2°, increased this to 0 . 0 02 . With absorption samples, this means that where virtually all of the wanted and none of the other polaris ation is absorbed, the ratio of light incident to light transmitted is . 0 02, giving maximum apparent absorbance 2 . 7 . This is over the instrument limit anyway, and weaker lines will not be affected .
In the Bausch and Lomb system, the dewar orientation
72 ducing an axial component into the o and TI spectra , again
in the ratio 0. 002 to 1 . In the present case , only very strong lines in TI will be affected.
4. 6 Temperature Measurement and Control
Two thermocouples were built into the dewar. For measurement near 4°K , a gold-cobalt versus copper thermo
couple with reference junction fastened to the bottom of the helium can was used. After the dewar had been in use for several months , this thermocouple began to give negative voltage readings , indicating poor thermal contact at the cold junction. This was subsequently found to be due to corrosion. Earlier temperature readings were therefore treated with care , and were usually rejected.
The second , more reliable thermocouple, was copper versus constantan , with an external reference junction.
As it could only be operated from a 77°K reference ,
sensitivity near helium temperatures was reduced , but still adequate. At 7°K the E. M. F. was -720 µV and the differential
2 µV/degree. Reading accuracy to 2 µV gave accuracy to 1°K , of the same order as control accuracy. At higher
temperatures accuracy was better than this .
In situ calibration of the thermocouples was not possible , so a standard temperature versus voltage table supplied with the dewar was assumed to apply. A small com puter program was used to convert the table to a more usable form with regular voltage increments.
Temperature control was manual only. Thermocouple voltage was read on an Andonian microvoltmeter. Any deviat ion from the desired temperature could be corrected by alter ing the power to the heater mentioned in section 4. 3. With practice, control was possible to within 1°K, which was quite adequate for the present work.
4. 7 The Complete Operating System
The complete assembly and associated services is shown with the Bausch and Lomb in Figure 4. 6. The layout was chosen as being the most convenient to operate and maintain.
The dewar on its stand could be raised out of the sample compartment and swung forward for sample changing and polar oid alignment. In the operating position, fine adjustment screws allowed precise location in the beam. Two-stage
mechanical pumps capable of producing pressures of 5 microns of mercury were used , one for pumping the vacuum space of the dewar, the other for pumping the heat exchange column, and the helium can when necessary. The outlet of this pump was
connected to the helium recovery system. Pump lines were supported so that the dewar was not under strain, and flex ible couplings were included to reduce vibrations.
Liquid nitrogen was transferred into the dewar by nit rogen gas pressure. Plastic tubing was adequate as transfer line. Liquid helium was transferred from storage dewars using an Andonian Associates transfer line. As 12 litres
74
were stored, and 6 were sufficient for one run, the transfer tube was only inserted part way into the full storage dewar and acted as a dipstick, stopping transfer when half the liquid had transferred. Sufficient helium was thus left to be sure of another run. Helium gas from the cylinder on the left was used for pressurising the heat exchange space and for purging the air from the dewar before liquid helium filling.
The vacuum gauge and temperature controller can be seen to the left of the dewar. This arrangement permitted temperature control and spectrometer operation from the same position.
Figure 4. 7 shows the same apparatus mounted on the Jarrell-Ash. Since the spectrometer was higher, operation was less convenient.