This section presents the results o f investigation o f the Eastman AQ29D/Immunodyne membrane combinations. Chronoamperometric measurements revealed weft defined transients (Figure 10.2 (a)) whereby we see a rise in the current response followed by a decay which reaches steady-state in under 10 s, after pulsing the voltage to +0.3 V. It is also apparent that the background current o f the gold microbands (with the Eastman polymer film on the surface) is low and stable. The measurements were repeated to show the reproducibihty o f the response (Figure 10.2(b)). Past studies (50) have shown that such an overlayer of Eastman polymer on the electrode wUl improve the linearity o f the glucose response and increases the dynamic range. The explanation given is that the film retains the redox couple on the surface of the electrode very well. The electrode-membrane system employed in this study was shown to be usable for the determination o f glucose concentrations up to 10 mM. As expected in each case, an

increase glucose concentration is accompanied by an increase in steady-state current. Once again, this increase in current response is attributed to the increase in Fe(CN6) production generated from the GOx reaction.

Cahbration o f the glucose response was observed by plotting the steady-state current response vs. glucose concentrations, as shown in Figure 10.3. The plots show a good linear response, with correlation coefficients of 0.96 and 0.98.

It has been estabhshed previously (Chapter 9.2.2) that the steady state current under excess mediator concentration is governed by the following equation:

/ = [glucose]

nk^

where L = v and = cothf ) = 1.0314 For the condition o f the

c o s h - g + l]

experiment AE = 0.3 V, ^ 2 = 1.014.

D = difrusion coefficient of mediator.

This equation predicts that plotting I vs. [glucose] will give a straight line whose slope allows us to evaluate the difrusion coefficient o f the mediator in the membrane films. From the cahbration plots (Figure 10.3), the difrusion coefficients determined are 2.55 x 10'^^ and 2.17 x 10^^ m^/s. Values are within two orders of magnitude lower than reported in bulk solution o f D = 6.5 x 10'^° m^/s. This is expected, as it is indicative of the effect of employing the membrane in the system.

Reproducibihty of membrane thickness was not a problem. Not only did the membrane allow us to have reproducibihty in the immobilization technique but also had the advantage o f aUowing high concentrations o f GOx and redox mediator without leakage of these species. In addition, the technique appears ideal for enzyme stabilization. The only disadvantage is that, prior to protein immobilization, the ‘pre-activated surface’ is unstable and slowly reacts with moisture in the air. Furthermore, although the membrane is known to have successfiiUy immobilized the GOx enzyme, the long-term stabihty o f the GOx is stiU questionable. A noticeable loss of activity was observed after 3 days, wherein results show a decrease in current response [Figure 10.4]. This may be due to the natural dénaturation o f enzyme with time.

During the transient measurements, the microbands were in contact with the reacting layer for 10 minutes at open circuit prior to applying a voltage o f +0.3 V. It was assumed that this aUowed the glucose to be ftiUy converted as weU as aUowing the

diffusion of Fe(CN6/' to take place. In addition, because the solutions were thoroughly degassed before use and the mediator was present in excess, it was assumed that the mediated signal had no oxygen dependency. In the next stage of our investigation, there was an interest to study the factors that control product detection and to see the effect of oxygen on the mediated signal. If the response is enzymatically controlled, then when glucose is added to the membrane, the shapes of the resulting transient will be governed by the kinetics o f the enzyme reaction. Results [Figure 10.5] show that there is generally an induction time before the current density increased from the baseline. The induction time seemed dependent upon glucose concentration, but not in any systematic way. A wide variety o f transient shapes were observed. The transients may have some artifacts due to the membrane drying up through time and therefore influencing the transport of Fe(CN6)'^' through the membrane. Overall, the response time of current should be linked to the distance through which redox species dijffuse through the membranes before reaching the electrodes as well as to the rate at which these species are produced. In our system, a combination o f membranes was employed. The diffusion of the mediator through the Immunodyne membrane is not the only possible limiting phenomenon. Diffusion through the polymeric Eastman film could limit the mediator flux reaching the electrode surface. Figure 10.6 shows the cahbration plot of final current obtained against glucose concentrations associated with these transients. The plot also shows the effect of having oxygen present. The ‘inflection’ point in the cahbration curve might be the result of the depletion o f oxygen in the enzyme layer. At concentrations below this inflection, re­ oxidation of the enzyme could occur mainly via oxygen, while above the inflection concentration, oxygen could be depleted in the layer adjacent to the electrode so re­ oxidation could proceed mainly via the mediator route, leading to an increase in the current response o f the enzyme electrode (56). This observation is similar to the results discussed in Chapter 9.2.2.

10.3.3 Use of conformai maps to model the chronomaperometric response of a