• No se han encontrado resultados

El post-rock y sus otros: indie, canonicismo y Joe Carducci

Capítulo 1. La llegada del post-rock a España (1992-1996)

1.1. Las definiciones de Simon Reynolds

1.1.1. El post-rock y sus otros: indie, canonicismo y Joe Carducci

EQUIPMENT:

1. Controlled constant temperature bath with continuously stirred liquid.

2. Certified Standard Thermometer.

3. A continuity test circuit.

4. High voltage breakdown tester.

5. A constant current source.

Note: Calibration baths are generally suitable for use only in workshop. If required for use in the field, they should be used under cover and in a non hazardous environment.

PRE-CONDITIONING: Before commencing any tests check the following: -

ERECTION, COMMISSIONING, OPERATION AND MAINTENANCE MANUAL FOR TRANSFORMER &

REACTOR

Document no: PTRM1104 Page 121 of 205

2. The instrument is not damaged in any way and any such damage is rectified before testing.

3. The instrument is mounted correctly. (Refer 6.1) 3. The shipping stop (22 - Fig. 1) is removed.

OTI CALIBRATION: Place the sensing bulb of the temperature indicator under test in the well-stirred constant temperature bath. Allow 5 minutes settling time before taking readings. Check the reading near to but not less than minimum value of the indicator scale. For increasing values of temperature, repeat the procedure to obtain at least two further readings the highest value being near to but not greater than the maximum scale value of the indicator. The readings should agree with the standard thermometer within

± 1.5% f.s.d. (Maximum scale value). Otherwise following adjustments should be made.

ZERO ADJUSTMENT: Use this adjustment for correcting readings near minimum scale value. Referring to Fig. 1, firmly hold the tongue adjuster (50 by a 2BA spanner and slacken the lock nut (4) with a second 2BA spanner.

To move the indicting pointer (19) up or down the scale, turn the tongue adjuster (5) by a small amount anticlockwise or clockwise respectively. Re-tighten the lock nut holding firmly the tongue adjuster. Do not twist the measuring bellows while carrying out this adjustment as this may cause damage to the linkage/bellows.

SPAN ADJUSTMENT: Use this adjustment for correcting reading near the maximum scale value. Referring to Fig. 1, slacken the shoulder screw (7) half a turn but do not remove the same. Turn the range adjuster screw (8) anticlockwise or clockwise to increase or decrease respectively the sweep of the pointer. Make the adjustment in small increments and carefully. Half a turn in either direction may correspond to approximately 0.5° reading. After adjustment, retighten the shoulder screw.

Ensure that the lock nut (4) and shoulder screw (7) are properly tightened after the above adjustments. Repeat the tests in cold and hot bath until the accuracy limit of ± 1.5% f. s. d. is achieved.

ERECTION, COMMISSIONING, OPERATION AND MAINTENANCE MANUAL FOR TRANSFORMER &

REACTOR

Document no: PTRM1104 Page 122 of 205

WTI CALIBRATION: During Calibration, keep the instrument lid fixed in position. Place the sensing bulb in the continuously stirred oil bath, the temperature of which is maintained at the specified value. Allow 5 minutes setting time and apply the specified current to the bellows heater and calibrating shunt network. The current must remain switched on at specified value for at least 40 minutes before taking reading. To achieve the readings within the accuracy limits or to reset the temperature rise to a different value, the following adjustments should be made.

GRADIENT ADJUSTMENT: After opening the instrument lid, carefully remove the dial. Withdraw the calibrating shunt (17-Fig. 1) after removing the nut and retaining washer. One adjustable centre band and two fixed bands at the ends are provided in the shunt. One supply lead is connected to the centre band and the other lead to the lower band on the shunt. Slacken the screw in the retaining clip of the centre band. To reduce the temperature rise, move the centre band nearer to the lower band, to increase the temperature rise reverse the procedure. Retighten the screw on the centre band after adjustment.

Replace the shunt, dial and lid. Repeat the calibration, check after adjustments until the accuracy limits achieved.

SWITCH SETTING: The switch settings are adjustable as indicated below: - Indicator Scale Range Switch setting adjustable Switch differential (close to

to close between open) adjustable between 0 - 120° C 40° C and 120° C 6° C and 80° C

20 - 120° C 50° C and 120° C 6° C and 90° C 30 - 120° C 50° C and 120° C 6° C and 100° C 0 - 150° C 45° C and 150° C 7.5° C and 105° C 0 - 240° C 80° C and 240° C 12° C and 160° C

ERECTION, COMMISSIONING, OPERATION AND MAINTENANCE MANUAL FOR TRANSFORMER &

REACTOR

Document no: PTRM1104 Page 123 of 205

NOTE: Unless specified otherwise, the instruments are supplied with minimum switch differential.

Connect across the switch contacts the continuity test circuit and check the correct operation of the switches at their set points to make (close) and break (open) contacts. The switches should operate at set point within ± 2.5% of switch setting arm pointer (10-Fig. 1) indication, otherwise the following adjustment should be made.

SWITCH ADJUSTMENTS: Referring to Fig. 1 slacken switch setting screw (11) on each switch setting arm (10). Move the left hand arm pointer over the switch scale (9) to the required temperature for making (closing) contact and right hand arm pointer to the required temperature for breaking (opening) contact. Retighten both switch setting screws. By closing the left hand arm pointer to the right hand arm pointer minimum differential is obtained. A built-in stop automatically sets the mbuilt-inimum differential. While slackenbuilt-ing and tightening the setting screw, steady the disc with the other hand to prevent undue strain on the mechanism. The indicating pointer can be moved with a finger in a clockwise direction to check the operation of the switches at set points. By moving the pointer slowly and steadily, an accurate check is obtained. Never allow the pointer to spring back suddenly or quickly and also force the pointer in an anticlockwise direction. By doing so the measuring bellows is likely to be damaged. An alternative procedure is to check the switch operation while calibrating the instruments for temperature readings.

INSTALLATION

MOUNTING: Mount the instrument in a vertical position. Otherwise errors, particularly in the horizontal plane, will upset the zero setting of the mercury switches. Place a sprit level on the top of the indicator case to ensure correct positioning and operation of the mercury switches.

If vibration exists at the mounting position, mount the indicator on a non-vibrating structure. Suitable anti-vibration mountings are supplied. Do not use these mountings during transport.

CAPILLARY LINE: Do not stretch hard twist or bend the capillary to a radius less than 75 mm. Continuously support the capillary by suitable clips at intervals of 300 to 450 mm. Allow necessary slack at the bulb and for

ERECTION, COMMISSIONING, OPERATION AND MAINTENANCE MANUAL FOR TRANSFORMER &

REACTOR

Document no: PTRM1104 Page 124 of 205

mounting and removing the bulb without sharp bend at the joint. Wind in spiral (minimum dia. 150 mm) the possible excess length of capillary close to the indicator or the bulb.