9. TRATAMIENTO DE LOS PACIENTES CON CÁLCULOS URETERALES
9.5 Recomendaciones relativas al tratamiento del paciente índice
The gamma camera and data acquisition electronics used on our laboratory are
designed to the FastSPECT II specifications [34]. The gamma camera is connected to
multi-channel pulse-height analyzers to read and digitize the output from the 9 PMTs
depicted in figure 2.5. The pulse-height analyzers are connected to a large memory buffer
via an ethernet cable, where data is stored in list-mode format. The electronics are then
controlled using LabVIEW [55]. The pulse height analyzers, measure the height of the
pulses generated from each PMT. For a single event, 9 numbers representing the pulse
height of each PMT are collected in a temporary memory buffer. During acquisition a list
of all events including their acquisition time and pulse height are accumulated over time
then stored for post processing. Data collected in this manner is known as a “list-mode”
acquisition [9, 78]. Using list-mode data, the projection images are calculated from the
involves two tasks; position estimation and the rejection of scatter events in the
acquisition process.
Ideally, all the energy of a gamma ray photon would be converted directly to
visible light, where all the visible photons are emitted along the same trajectory and
location of the incident gamma photons. As illustrated in figure A.1 a gamma ray
interaction with the NaI crystal results in a spread of visible photons. The energy of an
incident gamma ray is therefore spread over a range of pulse heights corresponding to the
range in energies of the secondary visible photons. Plotting the histogram of pulse heights
collected from a NaI detector, a peak corresponds to the energy of the gamma emitter.
The “photopeak” obtained from emissions of 99mTc correspond to an energy of 140 keV.
Ideally, a 100 keV gamma ray photon will produce an average of visible
photons in a NaI crystal [86]. The geometry of the crystal and the conversion from visible
5000±p5000
Figure A.1: Frequency spectrum of 99mTc with a photopeak corresponding to 140 keV (adapted from 14-4 in [23]). 140 keV photopeak (99mTc) Count s FWHM Digital units Compton Valley Compton Plateau
photons to electrons in the PMT can effect how broad the photopeak is for a given NaI
detector. The Full Width at Half Max (FWHM) of the photo peak is a measure of the
efficiency of the detector for a given gamma emitter [86].
The path of the incident gamma ray may also be redirected through non-ideal
interactions with the NaI crystal prior to conversion to visible light. These interactions are
known as scatter events and make up the low energy events in the energy histogram.
Scattering of gamma ray photons within the range of energies used clinically are
attributed to Compton scattering and the photoelectric effect [23, 86], and these
interactions are illustrated in figure A.2. In Compton scattering, a gamma ray photon
interacts with a free electron in the NaI crystal. In these collisions, some energy from the
incident gamma ray is transferred to the electron. To conserve momentum, the gamma
photon’s energy is reduced, and the trajectory is redirected. Subsequent scintillation
events within the NaI crystal will result in pulse heights corresponding to an energy lower
than the initial energy of the incident photon. In the photoelectric effect, all of the gamma
ray’s energy is transferred to a free electron within the NaI crystal. The free electron is
deflected and repeatedly interacts with the NaI crystal. The deflection interactions with
the crystal generate multiple low energy visible photons until the electron’s kinetic
energy is absorbed.
Additional features in the energy histogram can be attributed to background
events. Interactions between the gamma photons and lead shielding/collimators produce
x-rays that can interact with the NaI crystal [23, 86]. Gamma photons that pass through
the NaI crystal can interact with other components of the detector, such as the light guide.
back towards the NaI crystal, generating visible photons that are reflected off the back of
the detector face. These events are known as “backscatter” events [86]. Both lead shield
interactions and backscatter events form characteristic photopeaks in the Compton
plateau. Additionally, scattering events may occur within the subject/object being imaged,
which will increase the frequency of scattering events in the detector’s energy spectrum.
The list-mode pulse heights from each PMT therefore contain events due to
scattering. In order to calculate a projection image from the list-mode data the event’s
position must be determined and events falling within the Compton plateau must be
rejected, as the location of these events will not be consistent with the collimator
projection model. However, rejecting events using the energy spectra, known as energy
windowing [22], is complicated by the fact that the pulse heights from each PMT are not
only sensitive to the energy of events but also position. Rather, as part of the camera’s
calibration, the response of each PMT is measured by physically placing a well-
✓
Compton Scattering Photoelectric Effect
e-
e -
e -
e-
e -
Figure A.2: Scattering in the NaI crystal: (a) Compton scattering, (b) photoelectric effect (figures adapted from [85]).
collimated point source on a 78 x 78 grid of points spaced 0.15 mm apart on the face of
the detector. For each grid point, data is acquired until 5000 events are counted. The
mean response for all 9 PMTs at each grid point, known as the Mean Detector Response
Function (MDRF), is used to generate a look-up-table (LUT) from which events obtained
during an acquisition can be compared, as pictured in figure A.3 for one of our
laboratory’s cameras.
The location on the 78 x 78 grid for a given event is estimated using a
maximum-likelihood position estimator and treating the MDRF as the mean estimate for
all locations [22, 9, 43]. Finally, a “likelihood window” [22, 43] procedure is used
to accept or reject the event. Using the MDRF, the likelihood of the event occurring at the
(x, y)
(x, y)
Figure A.3: In situ mean response of each of the 9 PMTs to a well collimated point source positioned on a 78 x 78 grid in front of the camera (based on figure 12.2 in [21]).
estimated location with an energy of 140 keV is calculated. If the likelihood is above a
fixed threshold the event is accepted. In this way, list-mode-data is sorted to a grid
location or pixel. The entire projection is then the 2D grid of pixels where the value of a
given pixel is the number of events assigned to its location. Using a high likelihood
threshold, likelihood windowing demonstrates a similar probability of accepting 140 keV
events to energy windowing, considered the gold standard, and Bayesian windowing
Appendix B
Maximum Likelihood Expectation Maximization