• No se han encontrado resultados

Mass  spectrometry  (MS)  is  a  technique  used  to  measure  the  mass  and  give   insights  into  the  structure  of  a  given  molecule.  Molecules  must  be  in  the  gas  phase   for   MS   analysis   and   the   must   be   ionised   (either   bofore   or   after   entering   the   gas   phase).  Usually  an  electric  field  is  applied  to  the  gaseous  ions  to  accelerate  them   and  they  are  then  separated  by  their  mass  to  charge  ratio  (m/z).  

There   are   many   types   of   mass   spectrometer   arising   from   the   different   combinations  of  ionisation  and  detection  methods.  Electron  impact  is  a  method  of   ionisation   for   small   volatile   molecules   that   uses   electrons   to   ‘knock   off’   other   electrons  from  the  molecule  of  interest,  creating  a  positively  charged  radical  that   may  undergo  further  transformations  in  the  gas  phase.  Chemical  ionisation  uses  a   reagent   gas   that   is   ionised   and   collided   with   the   analyte   transferring   the   charge.   Matrix-­‐assisted  laser  desorption  (MALDI)  uses  laser  light  to  vaporise  the  molecules   of  the  matrix  containing  the  analyte.  The  matrix  is  designed  to  strongly  absorb  the   light   of   the   laser,   and   it   can   then   transfer   its   energy   to   the   sample.   A   potential   difference  is  then  applied  to  the  plume  of  analyte  and  the  sample  becomes  ionised.   This  ionisation  technique  is  much  ‘softer’  than  those  previously  described,  and  can   be  used  for  large  molecules  such  as  proteins.84,155  

In  the  work  carried  out  in  this  thesis,  only  electrospray  ionisation  (ESI)  was   used.  ESI  is  a  very  soft  ionisation  technique  that  requires  a  liquid  sample  input  and   can  be  used  for  the  analysis  of  small  ions  up  to  large  proteins.  Sample  solution  is   flowed   through   a   capillary   to   the   tip   where   a   potential   difference   is   applied.   The   voltage   applied   partly   determines   the   charge   on   the   sample   ions.   The   solution   forms  charged  droplets  from  which,  with  the  aid  of  the  sheath  gas  (typically  N2),  the  

solvent  molecules  are  evaporated.  As  the  droplets  shrink,  the  desolvation  becomes   more  rapid  due  to  Coulombic  forces  that  eventually  overcome  the  cohesive  forces   and  the  analyte  is  released  from  the  droplet  solvent-­‐free.155  Figure  1.24  shows  an  

illustration  of  the  process.  

  Figure  1.24  An  illustration  of  the  electrospray  ionisation  process.156  

 

As   mentioned   above   different   methods   of   ionisation   can   be   coupled   with   different  mass  analysers,  and  either  used  on  their  own  or  in  combination  with  each   other.  The  following  mass  analysers  were  used  to  obtain  data  in  this  thesis.    

One   of   the   simplest   mass   analysers   is   a   quadrupole.   It   consists   of   four   parallel   rods   to   which   a   constant   voltage   and   an   oscillating   radio   frequency   are  

applied  to  opposite  pairs  of  rods.  Ions  pass  through  the  rods  and  their  direction  of   travel  is  controlled  by  the  frequencies  applied  to  the  rods.  Hence,  certain  ions  may   be  excluded,  depending  on  their  m/z  values.  

Time-­‐of-­‐flight   (TOF)   analysers,   as   the   name   suggests,   separate   ions   by   the   time   taken   for   them   to   travel   through   a   high   vacuum   drift   region   within   the   analyser   called   the   drift   region.   Ions   leave   the   ionisation   chamber,   and   several   thousands   of   times   a   second   a   voltage   is   applied   adjacent   to   the   sample   stream,   sending  pulses  of  ions  of  the  same  kinetic  energy  into  the  drift  region.  The  ions  are   then   separated   by   their   mass-­‐to-­‐charge   ratios.   To   improve   the   resolution,   more   advanced  instruments  use  a  reflectron;  this  is  a  class  of  ion  mirror  that  consists  of   electrodes   with   progressively   higher   charges   applied   to   them.   This,   in   effect,   doubles  the  drift  region  and  the  resolving  power  of  the  analyser  as  well  as  focusing   ions  of  the  same  m/z.  

An  ion-­‐tap  analyser  shares  some  similarities  with  the  quadrupole  analyser.   Radio  frequencies  and  applied  voltages  are  used  to  select  the  ions.  They  are  then   ‘trapped’  within  the  analyser  and  can  accumulate  before  they  are  expelled  towards   the  detector.  

Tandem   mass   spectrometry   (MS/MS)   is   another   technique   used   in   the   course   of   this   work.   This   term   can   be   used   to   describe   a   number   of   ion   fragmentation  techniques,  but  in  this  case  it  refers  to  collision-­‐induced  dissociation   (CID).   For   this   process   an   ion   of   interest   in   selected   and   isolated,   often   using   a   quadrupole   or   an   ion   trap.   An   inert   gas   (in   this   case   N2)   is   then   introduced   and  

In   terms   of   application   to   the   study   of   metal   complexes,   high-­‐resolution   mass  spectrometry  has  been  used  for  the  characterisation  of  complexes,  to  study   their   interaction   with   small   biomolecules   such   as   GSH,157   their   interactions   with  

proteins,158,159  and  their  interactions  with  DNA.160–164    

Aims  of  this  work  

In   order   to   understand   the   mechanisms   of   action   of   metallodrugs,   it   is   important  to  develop  new  methods  for  their  characterisation,  and  in  particular  their   speciation  in  solution.  It  is  not  only  important  to  identify  target  sites,  but  also  to   determine  the  number  and  types  of  ligands  which  are  bound  to  the  metal  at  the   target  site,  and  hence  to  determine  how  the  initial  ligands  in  the  metallodrugs  are   involved  in  activation  and/or  target  reconition.  

 

The  specific  aims  were  as  follows.  

1. To   use   HPLC   to   characterise   the   lipophilicity   of   a   series   of   photoactivatable   Pt(IV)   anticancer   complexes   for   which   partition   coefficients  could  not  readily  be  determined  by  conventional  octanol-­‐water   shake   flask   methods,   and   to   attempt   to   correlate   their   lipophilicity   with   cancer  cell  uptake  and  biological  activity.  

2. To  explore  the  use  of  chiral  HPLC  columns  to  separate  enantiomers   of  organometallic  anticancer  complexes,  especially  Ru(II)  and  Os(II)  arenes,   and   cyclopentadienyl   Ir(III)   complexes.   Such   separation   is   important   for  

biological   testing   and   target   site   recognition,   which   can   show   a   strong   dependence  of  the  chirality  of  drugs.  In  particular,  the  aim  was  to  examine   the  stability  of  separated  enantiomers  and  complexes  with  subtle  difference   in  structure  such  as  facial  chirality.  

3. To   explore   the   use   of   new   methods   involving   the   coupling   of   photonic  crystal  fibres  with  high-­‐resolution  mass  spectrometry  that  may  be   more  efficient  both  in  terms  of  the  volume  of  analyte  required  (nanolitres),   in   terms   of   the   extent   of   photochemical   conversion,   and   the   detection   of   short-­‐lived  photoproducts.                              

Documento similar