Prototype conventional Linacs have been developed to run in a mode where the flattening filter is not present in the beamline. The beam parameters of these Linacs have been published by a number of authors from different sites (Ponisch et al., 2006; Hrbacek et al., 2010; Kry et al., 2007a; Vassiliev et al., 2006b; Stathakis et al., 2009).
The backscatter to the ionisation chambers was found to be comparable with the FFF beam compared to the cFF (Zhu et al., 2006). This confirms the results of previous Monte Carlo studies (Titt et al., 2006c). The dose-rates measured for FFF beams were reported to increase by a factor of 1.99 (Stathakis et al., 2009) to 2.26 (Hrbacek et al., 2010) compared to cFF beams at 6MV and 3.62 (Stathakis et al., 2009) to 4.03 (Hrbacek et al., 2010) compared to cFF beams at 10MV. As with the reported increase in doserate observed from Monte Carlo studies this is expected as the flattening filter acts as an attenuator in the beam reducing the doserate.
The in-air output ratio changes less with field-size for FFF beams compared to cFF beams (Zhu et al., 2006), this is expected as the flattening filter is the dominant source of scatter within the treatment head with head scatter as the dominant cause of in-air output ratio changes with respect to field size. The surface dose for FFF beams was measured to be 10% higher than for cFF beams. Also, the depth of maximum dose (dmax) was closer to the patient surface for FFF beams (Hrbacek et al., 2010; Vassiliev et al., 2006b; Stathakis et al., 2009). The increased surface and shallower dmax is a result of the higher proportion of low energy photons in the FFF beam. The variation of the depth of maximum dose varies less with field size for FFF beams so that a 40cm x 40cm field has a depth of maximum dose deeper than that for a cFF beam (Hrbacek et al., 2010; Vassiliev et al., 2006b). This is due to the reduction in head-scatter associated with FFF beams. The profiles of FFF beams are closely matched to cFF beams for field sizes smaller than 4cm x 4cm. The measured PDDs for a 6MV FFF beam more closely resembled that of a 4MV cFF beam with the 18MV profile being more closely matched to that of a 15MV cFF beam (Vassiliev et al., 2006b). The penumbrae for 6MV FFF beams were 0.3mm sharper than cFF for depths up to 12cm but for depths over that the cFF beams had a sharper penumbra when compared to FFF. The penumbrae for 10MV FFF beams were sharper for all depths with the largest difference of 0.6mm at the depth of maximum dose (Hrbacek et al., 2010). Other authors stated a reduction of penumbrae of 0.6mm for a 6MV beam and 0.9mm with an 18MV beam (Ponisch et al., 2006).
21
The collimator exchange effect describes how the output of an Xcm x Ycm treatment field differs from a Ycm x Xcm field, due to the different location in the treatment head of the X and Y diaphragms. The effect is reduced in FFF beams, being 1.8% for cFF beams at 6MV and 1.2% for FFF beams. The values are 2.4% for cFF and 1.5% for FFF at 10MV (Hrbacek et al., 2010). This is likely a result of the reduction in head scatter for FFF beams.
The measured leakage through the MLC from an FFF beam has been reported as being reduced compared to cFF beams: by 0.29% at 6MV and 0.16% at 18MV (Ponisch et al., 2006). This reduction is due to the softer energy spectrum of the FFF beams.
Neutron production in high Z materials within the linear accelerator head can contribute to both patient and operator dose. The flattening filter is a significant source of neutron production and the effect of the removal of the flattening filter on the number of neutrons produced in an 18MV photon beam has been studied (Kry et al., 2007a). The number of neutrons produced for each MU was 20% lower with the flattening filter removed.
Lang et al. (2012b) performed a study on the effects of FFF beams on the collection efficiency of ionisation chambers used to measure dose. The collection efficiency was determined using a two voltage technique (Lillicrap et al., 1990). The change in collection efficiency with respect to dose per pulse changed by 0.8% from the highest dose per pulse measured (1.7mGy per pulse) to the lowest dose per pulse (0.4mGy per pulse), the conclusion was that the tested air vented ionisation chambers could be used for relative dosimetry without the need for collection efficiency to be accounted for. Measurements on matched 6 and 10 MV FFF beams at St James’s University Hospital, Leeds yield a change in collection efficiency of 0.3% from 10cm deep to 20cm deep in water for 6MV FFF and 0.6% for 10MV FFF. The impact on the measured TPR results in a change of dosimetric calibration factor of 0% for 6MV FFF beams and <0.1% for 10MV FFF beams.
Many of the quality assurance parameters that are used for cFF beams such as beam flatness and current penumbra definitions are not appropriate for FFF beams, mainly due to the normalisation of the beams to the central axis as part of the definition. Fogliata et al. (2012b) proposed new normalisation and parameters such as unflatness and in-profile slope to characterise the FFF beams. The normalisation parameters were field-size dependent and were determined through measurements on beam profiles of unmatched FFF beams (Fogliata et al., 2012b). At this stage no proposal for clinically relevant tolerances were made by the author.
22
Further studies into the quality assurance of FFF beams on Varian accelerators have been performed by Clivio et al. (2014). These studies compiled data from 3 different TrueBeam machines that used the analysis parameters proposed by Fogliata et al. (2012b) to determine the stability of FFF beams over a year and determine reference levels for the QA of these FFF beams. The measurements were performed with ion chambers, the Starcheck (PTW,Freiberg) detector array and the electronic portal image detector (EPID) panel. The findings of the study suggest that the proposed slope parameter is susceptible to the positioning of the measurement device. The peak position parameter was also susceptible to variation depending on the resolution of the measuring device. Based on this study the tolerances for these measurement parameters were found to be a change from baseline of 0.04 for the unflatness value, 0.1%/mm for the slope and 1mm for the peak position. This investigation has not been performed on ‘matched’ FFF beams such as those produced by the VersaHD linear accelerator (Elekta AB, Sweden). Further work by Fogliata et al. (2016b) confirmed that the normalisation parameters defined in the earlier work were consistent across two additional TrueBeam Linacs. The work did conclude that the slope parameter previously described was susceptible to large variations, on average 6.3% between the different machines. The suggestion was made that the parameter be used as part of the assessment of peak position only.