Direct control
If a queue is still active, Direct control buttons will not be active. Make sure to pause the queue before opening Direct control / Service.
Action Step
In the menu bar click Direct Control.
1
Click the arrow below the system picture and select Direct Control.
2
Connect Alias autosampler to ÄKTA pure or ÄKTA avant Instructions 29040427 AC 55
9 Direct control and maintenance
Action Step
3
In the Alias - Direct control screen, the sample cooling, sample tray and washing can be selected.
Sample cooling: Sample cooling option (if installed) can be turned on/off. It displays the actual sample temperature and set point.
Sample tray: The sample tray can be moved to the front, to load the sample racks or micro plates and the tray can be moved to the rear position.
Tip:
By opening/closing the door of the Alias, the same functionality will be achieved.
Washing: By clicking the Initial wash button, the Initial wash will be started and will rinse the needle, the buffer tubing and the syringe. If the second wash port is activated in the settings, the second wash solvent can be also used to rinse the needle, the buffer tubing and the syringe.
Tip:
Perform an initial wash manually before starting a method.
SSV priming: In case of an installed Solvent Selection Valve, the SSV priming menu can be used to prime all lines.
The Initialize button can be used to initialize the autosampler, where all the modules will be set into the home position. It can also be used to reset an error state.
56 Connect Alias autosampler to ÄKTA pure or ÄKTA avant Instructions 29040427 AC
9 Direct control and maintenance
Action Step
In Alias - Direct control screen, click System info.
4
The System info screen is displaying the autosampler hardware information, like firmware version and serialnumber.
This information might be requested by the technical support department of the manu-facturer in case of an upgrade or service action.
Connect Alias autosampler to ÄKTA pure or ÄKTA avant Instructions 29040427 AC 57
9 Direct control and maintenance
Maintenance
Action Step
Select Maintenance.
1 2
In the Alias Maintenance control, there are four more screens available for controlling the autosampler.
I/O screen:
3
In the screen, the Auxiliary #1 can be switched On or Off.
This function can be used to check if the hardware connection to another device is properly made.
The integrated compressor can be switched On or Off, by selecting the PASA Compressor.
58 Connect Alias autosampler to ÄKTA pure or ÄKTA avant Instructions 29040427 AC
9 Direct control and maintenance
Action Step
Valves screen:
4
The screen, displays the position of the Injection valve, syringe valve and the optional Solvent Selection Valve.
Positions of the valves can be changed and will be displayed accordingly.
Connect Alias autosampler to ÄKTA pure or ÄKTA avant Instructions 29040427 AC 59
9 Direct control and maintenance
Action Step
Log counters screen:
5
The autosampler Log counters screen provides the details on the movements of the In-jector valve, the Syringe valve and the Syringe.
A service technician can reset if one of the modules have been replaced.
60 Connect Alias autosampler to ÄKTA pure or ÄKTA avant Instructions 29040427 AC
9 Direct control and maintenance
Action Step
Syringe / needle screen:
6
• Syringe and needle can be moved to Home, Exchange and End positions
• Related positions of needle and tray can be adjusted.
• Syringe can be adjusted.
Connect Alias autosampler to ÄKTA pure or ÄKTA avant Instructions 29040427 AC 61
9 Direct control and maintenance
Action Step
System info:
7
Displays information on Instrument firmware and used Driver.
62 Connect Alias autosampler to ÄKTA pure or ÄKTA avant Instructions 29040427 AC
9 Direct control and maintenance
GE, GE monogram , ÄKTA and UNICORN are trademarks of General Electric Company.
Alias is a trademark of Spark Holland BV. Excel, Microsoft and Windows are registered trademarks of Microsoft Corporation.
Images of the Alias autosampler and dialogs from the SparkLink software are reproduced with permission from Spark Holland BV.
The SparkLink software and the Alias autosampler are non GE Healthcare products manufactured and sold by third parties and GE Healthcare assumes no liability whatsoever for the performance or use of these products.
Any use of UNICORN is subject to GE Healthcare Standard Software End-User License Agreement for Life Sciences Software Products. A copy of this Standard Software End-User License Agreement is available on request.
UNICORN 6 © 2009-2016 General Electric Company
© 2013-2016 General Electric Company First published Apr. 2013
All goods and services are sold subject to the terms and conditions of sale of the company within GE Healthcare which supplies them. A copy of these terms and conditions is available on request. Contact your local GE Healthcare representative for the most current information.
GE Healthcare Europe GmbH
Munzinger Strasse 5, D-79111 Freiburg, Germany GE Healthcare UK Limited
Amersham Place, Little Chalfont, Buckinghamshire, HP7 9NA, UK GE Healthcare Bio-Sciences Corp.
100 Results Way, Marlborough, MA 01752, USA GE Healthcare Dharmacon, Inc.
2650 Crescent Dr., Lafayette, CO 80026, USA HyClone Laboratories, Inc.
925 W 1800 S, Logan, UT 84321, USA GE Healthcare Japan Corporation
Sanken Bldg. 3-25-1, Hyakunincho Shinjuku-ku, Tokyo 169-0073, Japan
For local office contact information, visit www.gelifesciences.com/contact GE Healthcare Bio-Sciences AB Björkgatan 30
751 84 Uppsala Sweden
www.gelifesciences.com/AKTA
29040427 AC 05/2016 a346