PUBLICACIONES DE PROGRAMAS DE DESARROLLO Y SU ANCLAJE EN LOS TERRITORIOS
LOS SEGUROS AGRÍCOLAS Y SUS EFECTOS EN EL AGRO DE ESTADOS UNIDOS
Diclofenac was extracted from water samples using a filter absorption method as
described in detail in Chapter 3. Water samples of 100 mL volume were transferred to Schott
bottles, acidified to a pH of 2 using approximately 3 drops of H2SO4, and had a Sartorius
Biolab 4.8 cm diameter, 0.2 µm nylon membrane filter added. The sample was then covered
in aluminium foil to protect against photo-degradation and put on a shaker table for 24 hours
at medium speed. The filter was removed and transferred to a 15 mL amber vial, which was
then placed on a heating block at 65-100°C to dry. The filter was rinsed three times using 4
mL of MeOH, and was vortexed during every rinse to ensure maximum extraction of
diclofenac from the filter and the aliquots of MeOH were combined in a 15 mL amber glass
vial. The MeOH was then evaporated off under nitrogen at 65°C until dry. The vial was
rinsed three times with MeOH (500 µL, 250 µL, 250 µL) and the aliquots of MeOH were
combined in a 15 mL amber glass vial..
Each extraction batch (n = 24 samples) included a spiked sample and a Milli-Q blank.
Two spike concentrations (0.1 and 200 µg/L) were used dependent on the expected sample
concentration.
2.6.2. High performance liquid chromatography analysis
HPLC is an analytical method that separates compounds dissolved in a solvent, based
on their structure and size due to their interactions with a solid adsorbent column (Kupiec,
2004). HPLC was used in this project as it is an affordable method of sample detection that
works well for high environmental concentrations of diclofenac (Nováková et al. 2006).
24
Samples were analysed on a Dionex UltiMate 3000 HPLC fitted with a photodiode
array (PDA) detector, quaternary pump, mobile phase degasser, and autosampler. The
column used was a Phenomenex Luna 5u C18 (2) column (150 x 4.60 mm). The mobile
phase was 60:40 NaH2PO4 buffer, pH 3: acetonitrile, run isocratically at 1 mL per minute.
Injection volume of the test solution was 20 µL with a total analysis time of 31 minutes per
sample. Ultra-violet PDA detection of diclofenac was at 240 nm with a retention time of 13.2
minutes.
A calibration curve of diclofenac standards was prepared for each run with a
concentration range of 0 to 2000 µg/mL. The calibration standards were run before and after
each batch of samples. The 100 and 1000 µg/mL standards, and a blank of AcN were
analysed after every 10 samples, while every 20 samples a sample was injected in duplicate
to monitor variability. These samples confirmed the stability and reproducibility of the
calibration. Calibration curves were linear (R2 = 0.9995) over the calibration concentration
range.
2.6.4. Gas chromatography–mass spectrometry analysis
GC-MS is an analytical method that combines two commonly-used techniques to
successfully separate volatile and semi-volatile compounds while also selectively detecting
them. Compounds are separated via GC, before being pumped into the MS where they are
ionised. Upon ionisation, the individual compounds are exposed to a beam of electrons that
induces fragmentation which results in a determination of mass that can be divided by the
charge (Fialkov et al. 2007). This allows determination of the molecular weight. Because of
this specificity, the GC-MS detects compounds at levels as low as sub-nanograms (Sneddon
et al. 2007). The GC-MS is a popular technique for qualitative and quantitative determination
25
2.6.5. Derivatisation
The samples prepared in Section 2.6.1 were evaporated until dry under nitrogen at
80°C and then had 100 µL of AcN added. Standards were made, ranging in concentration
from 0 to 100 µg/L. To both the standards and the samples, 25 µL of hydroxypyrene (internal
standard) and 100 µL of MSTFA (derivatisation agent) were added, before being vortexed
and placed on a heating block at 80°C for 1 hour (Farré et al. 2007).
2.6.6. Analysis of derivatised samples
The derivatised sample extracts and calibration standards were analysed by GC-MS
using a Shimadzu GC-2010 Gas Chromatograph, interfaced to a Shimadzu AOC-20i Auto
Injector and a Shimadzu GCMS-QP2010Plus detector. Instrumental control, data acquisition
and data processing were performed using the Shimadzu GCMS Solution software (Version
2.70). Analytes were separated on a Rxi-5Sil column (5% diphenyl/95% dimethyl
polysiloxane) 30 m x 0.25 mm ID, 0.25 µm film thickness, with an integrated guard column
(10 m, Integra-Guard) (Restek, Belleftone USA). Derivatised samples and calibration
standards were injected into the injection port in splitless mode at a temperature of 280°C in
1 µL volumes. The initial oven temperature of 80°C was held for 1 minute, then increased at
a rate of 10°C/min to 150°C, then increased at 8°C/min to 215°C, and then increased by
10°C/min until it reached the final temperature of 280°C. The total run time was 32.63
minutes. Helium was used as the carrier gas at a flow rate of 5.5 mL/min.
The ion source was held at 230°C and the GC-MS interface at 280°C. Electron Impact
Spectra (EIS) were obtained at 70 eV in selected ion mode (SIM). The MS was calibrated
against perfluorotributylamine (PFTBA) before each sample run using the mass spectrometry
26
Retention times and mass-to-charge ratios used for detection and quantification of the
individual compounds were as follows: diclofenac quantifying ion 214, qualifying ions 73,
367, and 352, retention time 20.91 minutes; hydroxypyrene (the internal standard)
quantifying ion 290, qualifying ions 275 and 259, retention time 22.65 minutes.
2.6.7. Quality assurance/quality control
To give quality control, controls were run with every exposure. This was to ensure
there was no contamination of the water used and to make sure the cleaning of the tanks was
being carried out properly. To provide quality assurance, exposure samples were run in
duplicate for the first exposure and in triplicate for the following two exposures. Blanks were
also carried out in duplicate or triplicate respective of the exposure week.
A water sample spiked with diclofenac was included in every batch to determine
method recovery. Water blanks were spiked with either 2 mL or 1 mL of diclofenac standard
(Section 2.4.1) depending on the expected concentration (200 µg/L and 0.1 µg/L
respectively). These samples then underwent the same extraction process as experimental
samples.
Duplicates of samples, high and low standards, and a blank were run during both
HPLC and GC-MS analyses to ensure stability and reproducibility of results. Standards were
run before the samples to ensure there was no interference with the readings such as early
elution or detection of the compound, or excess background noise. The blanks were used to
detect contamination from the machines and to act as a background reference when analysing
27