3.1 NOCIONES BÁSICAS DE LOS VALORES
3.2.5 Los valores en adolescentes
The formalism for the determination of absorbed dose to water in the msr field fmsr is detailed in Section 3.2.1. The measurement is performed using an ionization chamber with its reference point positioned at the reference depth zref in a water phantom8. Depending on the availability of a calibration coefficient
8 If a water equivalent plastic phantom is used, the expressions are modified according to the description given in Section 3.2.2.
for the ionization chamber, one of the three different methods outlined in Section 3.2.1 is used to determine the absorbed dose to water in a water phantom:
(a) A calibration coefficient, NDfmsr,w,Qmsr, in terms of absorbed dose to water for the ionization chamber in a reference beam of quality Qmsr in the msr field fmsr is available; this is the preferred option, although at present few standards laboratories offer this type of calibration. The absorbed dose to water for the fmsr field, in a beam of quality Qmsr, at the reference depth zref in water and in the absence of the ionization chamber is given by:
msr msr msr
msr msr msr
w,fQ = Qf Df,w,Q
D M N (24)
If the conventional reference field fref = 10 cm × 10 cm can be established at the machine, in Eq. (24) fmsr will be replaced by fref and the beam quality Qmsr by Q.
(b) A calibration coefficient, fref,w, 0 D Q
N , in terms of absorbed dose to water for the ionization chamber in a standards laboratory’s reference beam of quality Q0 in the conventional reference field fref = 10 cm × 10 cm is available, as well as a beam quality correction factor kQfmsr refmsr,,fQ0 to correct for the use of the calibration coefficient in the fmsr field. The absorbed dose to water for the fmsr field, in a beam of quality Qmsr, at the reference depth zref in water and in the absence of the ionization chamber is given by:
msr msr ref msr ref
msr msr 0 msr 0
, w,fQ = Qf Df,w,Q Qf ,fQ
D M N k (25)
When the calibration beam quality Q0 is 60Co, the generalized symbol for the beam quality correction factor kQfmsr refmsr,,fQ0 can be simplified to kQfmsrmsr,fref. If the conventional reference field fref = 10 cm × 10 cm can be established at the machine, in Eq. (25) fmsr will be replaced by fref and the beam quality Qmsr by Q. In addition, as the resulting double superscript fref in k can be removed (there is no need to consider different types of reference fields), Eq. (25) reduces to the formalism given in conventional COPs, showing consistency between those and the present COP.
While Eq. (25) is meant to be applied using kQfmsr refmsr,,fQ0 that are directly measured or calculated for the fmsr field, based on current knowledge and uncertainty estimates it can be assumed that kQfmsr refmsr,,fQ0 = kQ Qfref, 0, where Q refers to the hypothetical conventional reference field of the machine. Data for the beam quality correction factor kQfref, short notation for kQ Qfref, 0 with reference to Q0 = 60Co and fref = 10 cm × 10 cm, in WFF beams are given in Table 12 for a range of ionization chambers, consistent with the kQ data in Refs [1, 2, 7].
TABLE 12. kQfref DATA FOR THE CONVENTIONALfref FIELD (10 cm × 10 cm) FOR REFERENCE IONIZATION CHAMBERS IN WFF LINACS, AS A FUNCTION OF THE BEAM QUALITY INDICES TPR20,10(10) AND %dd(10,10)x Ion chamber TPR
20,10(10) = %dd(10,10)x = 0.630 63.4 0.660 65.2 0.690 67.6 0.720 70.5 0.750 73.9
Capintec PR-06C/G Farmer0.9970.9940.9910.9880.982 Exradin A2 Spokas0.9980.9970.9950.9920.988 Exradin A12 Farmer0.9980.9960.9930.9900.984 Exradin A12S0.9960.9940.9910.9870.981 Exradin A190.9960.9930.9900.9850.980 Nuclear Assoc 30-751 Farmer0.9960.9930.9900.9850.979 Nuclear Assoc 30-752 Farmer0.9970.9950.9920.9890.983 NE 2505/3, 3A Farmer0.9960.9940.9920.9890.984 NE 2571 Farmer0.9970.9940.9920.9890.984 NE 26110.9960.9930.9910.9880.984 PTW 23331 rigid0.9960.9920.9890.9850.980
↓
TABLE 12. kQfref DATA FOR THE CONVENTIONALfref FIELD (10 cm × 10 cm) FOR REFERENCE IONIZATION CHAMBERS IN WFF LINACS, AS A FUNCTION OF THE BEAM QUALITY INDICES TPR20,10(10) AND %dd(10,10)x (cont.) Ion chamber TPR20,10(10) = %dd(10,10)x = 0.630 63.4 0.660 65.2 0.690 67.6 0.720 70.5 0.750 73.9
PTW 23332 rigid0.9960.9930.9890.9840.978 PTW 23333 (3 mm cap)0.9960.9930.9890.9850.979 PTW 30001 Farmer0.9960.9930.9890.9850.979 PTW 30010 Farmer0.9960.9930.9890.9850.979 PTW 30002/30011 Farmer0.9960.9930.9910.9870.982 PTW 30004/30012 Farmer0.9980.9950.9930.9890.984 PTW 30006/30013 Farmer0.9960.9930.9890.9840.978 PTW 31003/31013 Semiflex0.9960.9930.9890.9840.978 SNC 100700-0 Farmer0.9970.9940.9910.9860.979 SNC 100700-1 Farmer0.9980.9960.9940.9900.984
↓
TABLE 12. kQfref DATA FOR THE CONVENTIONALfref FIELD (10 cm × 10 cm) FOR REFERENCE IONIZATION CHAMBERS IN WFF LINACS, AS A FUNCTION OF THE BEAM QUALITY INDICES TPR20,10(10) AND %dd(10,10)x (cont.) Ion chamber TPR20,10(10) = %dd(10,10)x = 0.630 63.4 0.660 65.2 0.690 67.6 0.720 70.5 0.750 73.9
Victoreen Radocon III 5550.9930.9890.9850.9790.973 Victoreen 30-3480.9950.9910.9880.9820.976 Victoreen 30-3510.9950.9910.9880.9830.977 Victoreen 30-3490.9950.9910.9880.9830.978 Victoreen 30-3610.9950.9910.9880.9830.977 IBA FC-65P (Wellhöfer IC 69) Farmer0.9970.9940.9920.9860.979 IBA FC-65G (Wellhöfer IC 70) Farmer0.9980.9970.9940.9890.983 Note: See Appendix I for details.
↓
If the calibration beam quality Q0 is not 60Co, the beam quality correction factor can be derived from the ratio of values for Q and Q0, as both are relative to 60Co. Hence, kQ Qfref, 0 is obtained from the values in Table 12 as follows:
ref ref
0 ref
0
, =
f
f Q
Q Q f
Q
k k
k (26)
(c) A calibration coefficient, fref,w, 0 D Q
N , in terms of absorbed dose to water for the ionization chamber in a standards laboratory’s reference beam of quality Q0 in the conventional reference field fref = 10 cm × 10 cm is available, but there is no beam quality correction factor available to correct for the use of the calibration coefficient in the fmsr field. The absorbed dose to water for the fmsr field, in a beam of quality Qmsr, at the reference depth zref in water and in the absence of the ionization chamber is given by:
msr msr ref ref msr ref
msr msr 0 0 msr
, w,fQ = Qf Df,w,Q Q Qf, Qf ,fQ
D M N k k (27)
This requires that the beam quality Q of the hypothetical conventional reference field of the machine (see Section 3.2.1.3) be estimated in order to adopt beam quality correction factors kQfref from Ref. [1] or from Refs [2, 7].
The determination of the beam quality index is detailed in Section 5.3.3, and the values of kQfref for a range of ionization chambers for generic WFF beams are given in Table 12. Based on current knowledge and uncertainty estimates, it can be assumed that kQfmsr refmsr,,fQ = 1, which is consistent with the assumption kQfmsr refmsr,,fQ0 = kQ Qfref, 0 in Eq. (25); note that these equalities hold only within the uncertainty estimates discussed in Appendix I.
In Eqs (24–27), MQfmsrmsr is the reading of the ionization chamber in the field fmsr corrected for influence quantities, such as pressure, temperature, incomplete charge collection, polarity effects, etc. (see Section 5.4).
5.3.2.2. FFF high energy X ray beams
For FFF high energy X ray beams, the formalism is essentially the same;
in particular, options (a) and (b) in the previous section remain unaltered. For the application of option (c), values of kQfref (the short notation for kQ Qfref, 0 when Q0 = 60Co) for a range of ionization chambers for FFF photon beams as well as
for the two specific machine types CyberKnife and TomoTherapy are given in Table 13. These data include a generic volume averaging correction factor9. 5.3.2.3. 60Co gamma ray beams
For 60Co gamma ray beams such as in the Gamma Knife, only option (b) is considered (see Eq. (25)). The factor kQfmsr refmsr,,fQ0 with reference to a chamber calibration fref,w, 0
D Q
N with Q0 = 60Co is close to unity for most chambers suitable for reference dosimetry in these treatment machines. Note, however, that as reference dosimetry in Gamma Knife beams is usually performed in plastic phantoms (ABS or Solid Water), the correction factors include the conversion to absorbed dose to water. Values of kQfmsr refmsr,,fQ0 for the Gamma Knife models Perfexion and 4C are given in Table 14.
5.3.3. Determination of the beam quality when the conventional fref cannot