• No se han encontrado resultados

Force and Torque Modeling for the Drilling of Bone for use in Orthopaedic Haptic Simulation Systems

N/A
N/A
Protected

Academic year: 2023

Share "Force and Torque Modeling for the Drilling of Bone for use in Orthopaedic Haptic Simulation Systems"

Copied!
139
0
0

Texto completo

92 Figure 5-19: A test comparing experimental validation data against the empirical model predicted force and torque using a feed rate of 2.0 mm/sec and a spindle speed of. The calibration and validation results of the empirical and neural network models for bovine bone drilling will be presented in Chapter 5.

Orthopaedic Haptic Simulation Systems

An overview of current orthopedic haptic simulation systems was presented, including the general requirements and limitations of systems currently available. Finally, the latest force and torque models used in current orthopedic haptic simulation systems were investigated.

Figure 2-1: Schematic diagram of an orthopaedic haptic simulation system displaying the virtual and real  world components
Figure 2-1: Schematic diagram of an orthopaedic haptic simulation system displaying the virtual and real world components

Geometric and Material Properties of Human Bone

Cortical Bone

To provide meaningful feedback, it is critical that the force prediction model providing the haptic response is accurate. The mechanical properties of elastic moduli, shear moduli, and Poisson's ratio of typical human cortical femur bone are presented in Tables 2-1, 2-2, and 2-3.

Cancellous Bone

Because cancellous bone is very heterogeneous, only certain mechanical and strength properties have been reported.

Mechanics of Drilling

In this image, a positive rake angle can be seen along the cutting lips towards the outer radius of the drill bit (view AA). As the cutting lips move toward the center of the drill, the rake angle becomes negative (view BB).

Figure 2-5: Tool-to-workpiece interaction that occurs during an orthogonal cutting process [27]
Figure 2-5: Tool-to-workpiece interaction that occurs during an orthogonal cutting process [27]

Drill Thrust Force and Torque Prediction Modeling

This suggests that the cutting forces and torques remain constant throughout the cutting of the hole. On the other hand, cutting brittle materials generates discontinuous chips that tend to clump together in the grooves of the drill.

Figure 2-7: a) Free body diagram of a drill tip displaying the tangential forces, b) Free body diagram of a  drill tip displaying the perpendicular and ploughing forces, and c) Area of material contact when cutting [11]
Figure 2-7: a) Free body diagram of a drill tip displaying the tangential forces, b) Free body diagram of a drill tip displaying the perpendicular and ploughing forces, and c) Area of material contact when cutting [11]

The Drilling of Bone

A selection of the results obtained is shown in Figure 2-11. From these figures, both the thrust force and the torque decrease significantly as the drill spindle speed increases. Experimental results showed a linear relationship between spindle speed and feed speed at two different locations of the bone (in the middle of the spindle and at the head of a femur bone) [31].

Figure 2-9: Experimental results relating feed with a) thrust pressure, b) torque, and c) specific cutting  energy using three different types of drill bits [9]
Figure 2-9: Experimental results relating feed with a) thrust pressure, b) torque, and c) specific cutting energy using three different types of drill bits [9]

Force and Torque Prediction Modelling for Haptic Rendering

Force and torque increase approximately linearly as the tip of the drill bit is engaging the material. Once the drill tip is fully engaged in the material, both force and torque level off and become constant for the remainder of the drilling operation. Once the drill tip is fully engaged, the cutting force and torque remain constant over the remainder of the drill.

Figure 2-15: Experimental results used to validate the proposed model at the exit of the cortical bone [12]
Figure 2-15: Experimental results used to validate the proposed model at the exit of the cortical bone [12]

Cutting Force and Torque Modeling

To provide an illustration of the effectiveness of these models, Figure 3-3 presents a comparative plot of the force formulation versus experimental data collected while drilling bovine bone. The obvious shortcoming of the model is not knowing the chip evacuation force due to the drill flute. To further emphasize this point, Figure 3-4 shows a clearer perspective of the same comparison this time showing a comparison of the force and torque experienced by the unicortical drill (albeit a single cortex).

Figure 3-3: Plot comparing Allotta
Figure 3-3: Plot comparing Allotta's force prediction model with raw experimental data collected drilling bovine bone

Chip-evacuation Force and Torque Modeling

Empirical Model

The frictional force Ffwi is due to the friction between the bone fragments and the wall of the hole. The frictional forces at the contact surfaces of the drill bit and the borehole wall will act in the opposite direction of the relative velocity. Chip removal torque is simply the product of the frictional force on the bore wall and the bit radius.

Figure 3-8: A cross-section of a typical twist drill bit.
Figure 3-8: A cross-section of a typical twist drill bit.

Artificial Neural Network

To properly substantiate the presented models, this section concludes with a list of the following assumptions: (1) the bone chips remain in constant contact with each other; (2) the evacuated material flows through and fills the drill groove; (3) the bone chips continuously exert pressure on the contact surfaces (i.e. the drill groove and the wall of the hole); (4) the helix angle of the drill is negligible; (5) the friction forces between the bone chips and their contact surfaces are proportional to their normal forces; and (6) the bone chips exhibit a non-isotropic stress distribution. The box with the f represents the transfer function chosen based on the application of the ANN. Typically, the amount of samples to be used for each training, validation, and test is determined based on a percentage of the total number of samples.

Figure 3-10: A single neuron artificial neural network.
Figure 3-10: A single neuron artificial neural network.

Finite Element Modeling

  • Human Femur Bone Model Acquisition
  • Boundary Conditions
  • Load Application
  • Meshing
  • Mesh Sensitivity Analysis

The second site of interest was at the proximal end of the bone (where the femur joins the hip). The final drill site examined is at the distal end of the femur (where the bone joins the knee). To perform a mesh sensitivity analysis, the femur was loaded with a realistic force and torque at the mid-diaphyseal point of the bone.

Figure 3-11: The three solid bodies used to form the model of a human femur bone; a) cancellous bone distal,  b) cancellous bone proximal, and c) cortical bone
Figure 3-11: The three solid bodies used to form the model of a human femur bone; a) cancellous bone distal, b) cancellous bone proximal, and c) cortical bone

Design of Experiments

  • Experiment Classification
  • Experimental Procedure
  • Model Parameter Definition
  • Model Calibration Procedure

When the drill penetrates the cortical bone and enters the cancellous bone, stop the feed of the CNC machine. As can be seen above, the force and torque prediction models are both a function of the operating parameters of the drill; namely, they are a function of the spindle speed and feed rate of the drill. This involved an inspection of the geometry of the drill to determine the contact lengths and the area of ​​the grooves of the drill.

Figure 4-1: The orthopaedic drill penetrating a cross-section of a bone a) bi-cortically, and b) unicortically
Figure 4-1: The orthopaedic drill penetrating a cross-section of a bone a) bi-cortically, and b) unicortically

Experimental Set-up

  • CNC Drill
  • Transducer and Data Acquisition
  • Drill Bit
  • Bone Specimen

Each of the drills used for experiments were unused and came directly from the manufacturer. A photo of one of the drills used for these experiments is shown in Figure 4-5. Figure 4-6 shows an image of one of the femur bone specimens loaded into the vise and ready for experimentation.

Figure 4-4: Screen shot of the Microsoft Excel data acquisition user interface.
Figure 4-4: Screen shot of the Microsoft Excel data acquisition user interface.

Drill Misalignment

This means that the axis of the bore path must be parallel to the z-axis of the sensor. Furthermore, these forces and torque would be due to the contributions of the cutting force and would actually reduce the force observed along the z-axis. Another misalignment problem that occurred was the misalignment of the drill path axis with the drill shaft axis.

Figure 4-8: An example of data collected for x-, y-, and z-axis forces.
Figure 4-8: An example of data collected for x-, y-, and z-axis forces.

Data Test for Normality

  • Motivation and Procedure
  • Data Collection
  • Median Rank Test
  • Chi Square Test

The result of the median ranking test for the initial cutting force is depicted in Figure 4-13. As with the test performed on the initial cutting force, this median rank test shows that the data are also trending. The Chi Square test is based on plotting the data in a histogram form.

Figure 4-13: The results of a median rank normality test using initial cutting force data
Figure 4-13: The results of a median rank normality test using initial cutting force data

Data Processing

Data Isolation

The second part of the data that is undesirable is the rapid increase in the cutting force and torque that is observed when the drill first engages the bone and continues until the initial cutting force is reached. This portion of the data is undesirable as prediction models for the initial cutting force are common in the literature and our model only attempted to predict the force and torque after the initial cutting force was reached. The last portion of the signal that is unwanted is the portion of data from when the drill breaks through the cortical bone until the data collection equipment is turned off.

Figure 4-17: An example of the raw a) force and b) torque data obtained experimentally
Figure 4-17: An example of the raw a) force and b) torque data obtained experimentally

Data Normalization

These three sections of data were removed manually, leaving only the drill clogging signal of interest.

Data Filtering

In addition, a low-pass filter will also remove any of the high-frequency noise inherent in the data acquisition system. Tuning the low-pass filter required an examination of the data to determine the filter parameters; namely: the sampling frequency (fs), the pass frequency (fpass) and the stop frequency (fstop). The filter parameters needed to properly tune the low-pass filter were strictly dependent on the experimental conditions; specifically on the conditions of the experiment which are time dependent.

Figure 4-20: Simulink schematic of the low-pass band filter used to filter experiential signal data
Figure 4-20: Simulink schematic of the low-pass band filter used to filter experiential signal data

Empirical Chip-Evacuation Model Calibration

Since jamming is a function of hole depth, experiments performed on thinner areas of bone lacked increased force and torque due to the jamming effect. This involved comparing the experimentally obtained values ​​with the empirical form of the force and torque prediction models using Eq. With the optimized friction coefficients determined for all data sets, the next step was to determine the power law coefficients.

Figure 5-1: A complete set of experimental data collected using a feed rate of 2.0mm/sec and a spindle speed  of 1500RPM
Figure 5-1: A complete set of experimental data collected using a feed rate of 2.0mm/sec and a spindle speed of 1500RPM

Empirical Chip-Evacuation Model Validation

Figures 5-3 and 5-4 show the thrust and torque using a feed rate of 1.75mm/sec and a spindle speed of 1250RPM for each of the two tests. Figures 5-5 and 5-6 show the thrust force and torque using a feed rate of 2.0mm/sec and a spindle speed of 1000RPM for each of the two trials. Also in each figure is the respective empirical model's prediction of the chip evacuation force and torque based on the operating parameters.

Table 5-3: The operating parameters used for validation experiments.
Table 5-3: The operating parameters used for validation experiments.

Empirical Chip-Evacuation Model Sensitivity Analysis

This is due to the force caused by μw acting perpendicular to the thrust and adding only a marginal effect to the total force. This can be explained by observing that the effect of μw occurs at the radius of the drill, where an increasing resistance at the contact between material accumulation in the groove and the wall of the hole is in the opposite direction to the rotation of the drill. . Furthermore, as discussed, the effect of μf acts perpendicular to the drilling surface and thus has a reduced effect on drilling torque.

Artificial Neural Network Training

Network Neuron Sensitivity Analysis

In general, increasing the number of neurons in the hidden layer increases the flexibility and processing power of the system. It is therefore important to determine the optimal number of hidden layer neurons that will be used by the network. The first step in this analysis was to select the range of hidden layer neurons to use to determine the sensitivity of the network.

Table 5-7: Operating parameters used for network sensitivity and validation.
Table 5-7: Operating parameters used for network sensitivity and validation.

Artificial Neural Network Validation

The newly obtained data were processed and plotted against the predicted chip evacuation force and grid torque output. As can be seen in Figures 5-14 to 5-17, the two-neuron artificial neural network predicts the chip evacuation thrust force and torque for drilling the bovine bone well. Based on the relatively high R-squared values ​​shown in Table 5-8, it can be concluded that the output obtained for the two-neuron artificial neural network is acceptable to estimate the increased thrust force and torque experienced during drilling of beef bone.

Table 5-8: The coefficients of determination, R2, for the force and torque network outputs for each of the  four validation experiments
Table 5-8: The coefficients of determination, R2, for the force and torque network outputs for each of the four validation experiments

Empirical Chip-Evacuation Model Calibration – Human Bone Drilling

As can be seen for Figures C-6 to C-10 of Appendix C, the experimental data collected for the chip evacuation force and torque for drilling human bone showed much greater variation than the data collected for the drill from beef bone. There are some trends that are similar to the trends seen in beef bone drilling. In addition, as seen when drilling beef bone, as the feed rate increased while the spindle speed remained constant, the coefficient of friction between the drill flute and the bone chip, μf, increased.

Table 5-9: Specification data for each of the 16 human femur bone specimen used for experimentation
Table 5-9: Specification data for each of the 16 human femur bone specimen used for experimentation

Empirical Chip-Evacuation Model Validation – Human Bone Drilling

As was done with the validation for drilling bovine bone, the accuracy of the results for drilling human bone can be measured by calculating the coefficient of determination. This discrepancy can be attributed to the variation in material properties of the bones used for calibration and validation. This greatly affects the consistency of the compressive force, as softer bones would require less force, while harder bones would require more force.

Table 5-13: The coefficients of determination, R 2 , for the force and torque models for each of the two  validation experiments
Table 5-13: The coefficients of determination, R 2 , for the force and torque models for each of the two validation experiments

Comparison of Results

Empirical Model versus Artificial Neural Network

Drilling of Bovine Bone versus Drilling of Human Bone

As mentioned previously, the relatively poor correlation for the human bone model validation and the large variation in the friction coefficient, μw, can be attributed to the inconsistencies in material properties of the human bone specimens. The donors of the femur bones used for calibration and validation experiments varied between men and women and had a wide range in age. For this reason, it can be concluded that although the presented models adequately predict the tension force and torque experience during drilling of human bone, experiments performed on specimens of the same sex and of similar age would have provided much more accuracy. Results.

Empirical Model Comparisons to the Literature

This resulted in a large disparity in bone strength and hardness between the samples, which was not present in the experiments where bovine bone was drilled.

Empirical Model Implementation into FEA Simulation

FEA Simulation Solution and Post-Processing

Also depicted in this figure is a plot showing the maximum deflection as a function of time. The results indicate that the maximum deflection experienced by the bone as a result of dynamic loading is more than 4 times the deflection resulting from static loading. The maximum deflections for the three locations using the two models are shown in Table 5-14.

Figure 5-23: The result of the FEA simulation using the presented force and torque models while drilling  mid-diaphysis
Figure 5-23: The result of the FEA simulation using the presented force and torque models while drilling mid-diaphysis

Summary of Contributions

Concluding Remarks

A comparison was made between the empirical model for drilling bovine bone with two models from the literature. These results were compared with those obtained using one of the static models from the literature. It was found that the deflection and equivalent stress were almost five times greater when the dynamic loading of the presented models was applied.

Future Work

It was concluded that the established models were acceptable for predicting the chip evacuation force and torque for drilling human bone, but more accurate results could be obtained if greater control over biological variations of specimens was used. This comparison showed that none of the formulations taken from the literature were able to accurately predict the experimental thrust force. Handbook of Fractures: 3rd Ed. evaluation of the less invasive stabilization system for the internal fixation of distal femur fractures.

Table B-1: The exponential constant, p, and the initial cutting force (F o ) for all normality test data
Table B-1: The exponential constant, p, and the initial cutting force (F o ) for all normality test data

Figure

Figure 2-1: Schematic diagram of an orthopaedic haptic simulation system displaying the virtual and real  world components
Figure 2-9: Experimental results relating feed with a) thrust pressure, b) torque, and c) specific cutting  energy using three different types of drill bits [9]
Figure 2-13: Experimental results indicating the three way relationship between spindle speed, thrust force,  and power consumption [32]
Figure 2-14: Pictorial representation of formulation parameters; a) infinitesimally small bone layer thickness,  dx; b) material transition [12]
+7

Referencias

Documento similar

“la tragedia del afrancesado español” (RODRÍGUEZ CACHO, 2009, p. Los ilustrados españoles vivieron, en efecto, una compleja relación de amor y odio con Francia: por un lado