2. Desarrollo Experimental
2.2. Caracterización
The schematic diagram of experimental setup for evaluating the studied TSE
Systems twin-head electrospray (THES) disperser of nanoparticles is shown in Figure 4.1. The THES case is a metal 6-way cross (i.e., formed by three straight cylindrical channels
perpendicular to and intersecting with each other), thus having six openings. The top and the
54
of particle-laden flow, respectively. The volumetric flow rate of primary carrier air is
variable by a needle valve and monitored by a flow meter. The front and the back openings are used as windows to illuminate the spray chamber by a light source and to observe the
spray process by a microscopic lens and a CDD camera. The left- and right-hand side
openings of the spray chamber are used to install two identical electrospray capillaries,
aligned in the axial direction and faced against each other. The tested stainless steel
capillaries have an inner diameter (ID) of 0.7 mm and an outside diameter (OD) of 1.0 mm.
A sheath flow compartment is also included in the installation of each capillary to provide
clean air flow sheathing the capillary. To operate the THES disperser, a positive DC high
voltage was applied on one capillary and a negative voltage on the other (by the H.V. power
supplies: Bertan Series 230, Spellman High Voltage Electronics Corporation, Valhalla, NY).
The spray chamber was electrically grounded. The spray suspension was fed into two
capillary needles by two identical syringe pumps (NE-300, New Era Pump System Inc.
Farmingdale, NY). For simplicity in operation, the feeding flow rates of two capillaries were
set the same. The flow rate of capillary sheath gases for two capillaries were also kept the
55
Figure 4.1: Schematic diagram of the studied twin-head electrospray disperser and its experimental setup for performance evaluation.
Distinguished from the typical cone-jet mode electrospray (applied in the majority
of literature), the THES disperser operated the spray at the stable multiple-jet mode to produce a stream of particles in high aerosol concentration. The sizes of nanoparticles to be
tested are pre-determined by their synthesis. There was no concern about altering the sizes
of nanomaterials. The multiple-jet electrospray used in the disperser is just a means to
disperse nanoparticles in air with the minimal agglomeration. Notice that the stable multi-
jet electrospray described herein was operated at the mode when the maximal number of stable jets at the capillary tip was observed by a CCD camera with a microscopic lens.
To investigate the performance of the studied THES disperser, a Scanning Mobility
Particle Sizer (SMPS) (including TSI 3080 electrostatic classifier and TSI 3025 UCPC, TSI
56
dispersed nanoparticles. During the operation, a camera with a microscopic lens was set up
to observe and record the liquid meniscuses at the capillary tips.
To characterize the mass concentration of dispersed nanoparticles in the carrier gas
stream output from the THES disperser, a stainless steel filter holder with a pre-weighted
HEPA filter (Whatman 934-AHT M, 47 mm φ, GE Healthcare Bio-Sciences Corporation,
Psicataway, NJ) was installed at the downstream of the THES to collect all the particles in
the gas stream exiting the system. The filter after collecting particles was kept in well-
ventilated lab environment (usually kept at 30% RH) for more than three hours prior to its
mass measurement to minimize the solvent contribution. The mass of particles collected on
the HEPA filter was calculated as the filter mass difference before and after the collection.
The total particle concentration in the exiting gas stream was derived as follows:
𝐶 (𝑚𝑔 𝑚3) =
𝑀𝑙𝑜𝑎𝑑−𝑀𝑐𝑙𝑒𝑎𝑛
(𝑄𝑀+2∗𝑄𝑆)𝑡 (4-1)
where 𝑀𝑐𝑙𝑒𝑎𝑛 and 𝑀𝑙𝑜𝑎𝑑 are the filter masses before and after particle collection, respectively; 𝑄𝑀 and 𝑄𝑆 are the flow rates of primary carrier and capillary sheath flows, respectively; t is the particle collection time in each run.
To achieve high mass concentration of particles in the gas stream, the following
parameters were varied during the study: needle tip distance, main carrier flow rate, side
sheath flow rate, liquid feeding rate and nanoparticle concentration in spray suspensions.
Three different nanomaterials, i.e., TiO2 (99.7%, <25nm, Sigma-Aldrich Co.), ZnO
(99.9+%, 20 nm, Nanostructured & Amorphous Materials Inc.) and NiO (20 nm,
57
three most popular nanoparticles reported in toxicity studies and applications. Previous
research showed that the breakage induced by nano TiO2 and the oxidative stress under the exposure of nano TiO2 could cause adverse effects on erythrocyte, implying a potential
toxicity to human health (Li et al., 2008). Brunner et al. (2006) found that nano ZnO with
the concentration above 15 ppm could kill almost all human or rodent cells. NiO was
classified as a Class 1 carcinogenic material by the International Agency for Research on
Cancer (IARC; IARC Monographs on the Evaluation of Carcinogenic Risk to Humans,
Volume 49 Chromium, Nickel and Welding 1990; cited in Horie et al., 2012). In this study,
spray liquids were prepared by dispersing nanoparticles (i.e., TiO2, ZnO and NiO) in
isopropanol (IPA)/aqueous (1:1 in v/v) mixtures at various concentrations ranging from 1.0
g/L to 10.0 g/L. A trace amount of nitric acid was added into the nanoparticle suspensions
to control the electrical conductivity of spray suspensions at approximately 150 µS/cm. To
keep the colloidal suspensions stable during the entire experimental run, the spray
suspensions were sonicated by an ultrasonic processor (CPX 750, Cole-Parmer Instrument,
IL).