The following conditions can prevent a real-time sensor glucose measurement from being taken: • A Lost Sensor alert.

A Sensor Error alert.

A new sensor that you just inserted is being initialized.A Cal Error alert.

More than 12 hours has passed since the last system calibration.Find Lost Sensor (15 minutes).

Weak signal alert.

Reconnect old sensor (two hours).

Opening and viewing the graphs

The following table describes how to use the graphs.

If you want to: Do the following:

Open the graphs From the HOME screen, press ESC. The 3-hour graph shows the details

for your most recent sensor glucose (SG) measurement. Press ESC again to open the 24-hour graph.

View information in a graph If there is no SG measurement at the time you select, the reason for

the missing measurement will be shown on the right side of the graph (for example, LOST SENSOR).

• When you open the graphs, the cursor (flashing vertical line) is always on the right edge of the graph and the graph shows the most recent SG measurement.

Press to move the cursor left to see an earlier SG measurement.Press to move the cursor right to return to more recent SG

The graphs

All graphs show High and Low Glucose lines, a real-time SG measurement line, a data section, and the cursor (flashing vertical line).

When you open any graph, the cursor flashes on the right edge of the graph. The data section shows the most recent SG measurement or the reason why no measurement shows, and the time in the data section matches the time at the top of the screen.

Indicates that your glucose has risen above 310 mg/dL (17.2 mmol/L). Sensor glucose data

points.

Each time you bolus, a marker will appear on your graph.

Cursor (flashing line) indicates selected time, and the sensor glucose measurement (or alert) for that time. There are three marks on the cursor at 100, 200, and 300 mg/dL (or 5, 10, and 15 if mmol/L is selected).

Data section shows the selected time, the type of graphs (or “History” if you have scrolled to the left on the graphs), and the sensor glucose measurement (or alert). High Glucose Limit Line at 191 mg/dL (10.6 mmol/L) Low Glucose Limit Line at 58 mg/dL (3.2 mmol/L)

3:16P

281

3 Hour

When you move the cursor left to select an earlier SG measurement, the data section of the graph turns black, the word History appears here, and the time in the data section changes to show the time when the SG measurement was taken or the reason why no SG measurement shows for that time.

2:20P History

LOST SENSOR

PM

NOTE:

There are many reasons your graphs may not show an SG measurement, but your system does

not alert for every one. Therefore, your system’s alarm/alert history may not match the number of times your graphs do not show an SG measurement.

Examples of real-time sensor glucose graphs

The following pages show examples of the different graph types.

3-hour graph

Each part of the graph between the vertical dashed lines represents one hour. In this example, the selected SG measurement was taken at 3:16 P.M., and the measurement at that time was 281.

3:16P

281

1:00 PM 2:00 PM 3:00 PM

3 hours

24-hour graph

Each part of the graph between the vertical dashed lines represents 12 hours. The shaded area helps you more easily see the sensor glucose measurements for the previous night. In this example, the selected SG measurement was taken at 11:50 A.M. and the measurement at that time was 211.

211

24 Hour

11:50A

The shaded area is the 12 hours from 6:00 PM to 6:00 AM

24 hours

How to check for rapid changes in sensor glucose

If your sensor glucose (SG) rises or falls faster than a certain rate, rapid change arrows automatically appear in the graphs next to the SG measurement, as shown in the following section.

The rapid change arrows

These examples show how the Rapid Change arrows tell you when your SG rises or falls faster than a certain per-minute rate. The arrow direction shows if your SG measurements are rising or falling. The number of arrows (one or two) shows how rapidly your SG is changing.

The arrow direction shows if your SG measurements are rising or falling. The number of arrows (one or two) shows how rapidly your SG is changing.

One up arrow shows that your SG has been rising at a rate of 1 to 2 mg/dL (0.05 to 0.11 mmol/L) per minute.

240

3:16P 3 Hour

One down arrow shows that your SG has been falling at a rate

of 1 to 2 mg/dL (0.05 to 0.11 mmol/L) per minute. 3:11P

131

3 Hour

Two up arrows show that your SG has been rising at a rate

of 2.0 mg/dL (0.11mmol/L) or more per minute. 3:11P

212

3 Hour

Two down arrows show that your SG has been falling at a

rate of 2.0 mg/dL (0.11 mmol/L) or more per minute. 3:11P

103

Calibration history

The SENSOR UPDATE HISTORY screen lists the successful sensor calibration BG values that were entered into the pump more than 15 minutes ago.

To view your Sensor Update History:

1 Go to the SENSOR UPDATE HISTORY screen.

Main > Sensor > Sensor Update Hist.

2 The SENSOR UPDATE HISTORY screen appears. It shows up to 28 calibration values.

Sensor alarm history

The SENSOR ALARM HISTORY screen lists all of the sensor alerts that have occurred, displaying up to 36 alerts.

To view your Sensor Alarm History:

1 Go to the SENSOR ALARM HISTORY screen.

Main > Sensor > Sensor Alarm History

2 The SENSOR ALARM HISTORY screen appears. The most recent alert is highlighted. Select the alert you

want to get more information on, then press ACT.

3 A new screen will appear with the alert details.

Disconnecting the transmitter and removing the sensor

Disconnecting the sensor from the transmitter

If you are not going to replace the sensor, turn the Sensor features Off to avoid getting a LOST SENSOR alert.

1

Hold the transmitter as shown, and pinch the flexible side arms of the sensor between your thumb and forefinger.

2 Gently pull the transmitter away from the sensor assembly. If you need to connect the transmitter to

the tester, wait one minute after disconnecting the transmitter from the sensor.

Removing the sensor

Gently pull the sensor from your body to remove it. Place it in a sharps container.

Storage

Using your system in water

WARNING: Do not use your pump in water or wear it during water activities. The pump is not water- tight and it may become damaged if it is used in water.

You should shower, bathe and swim with the transmitter and the sensor by following the guidelines below:

1 Disconnect the infusion set from the pump and remove the pump. The pump is not water-tight.

2 After the transmitter and sensor are connected, they form a water-tight seal to a depth of 8 feet (2.4

meters) for up to 30 minutes. You can shower and swim without removing them. Avoid taking hot baths as this may significantly reduce the sensor's life.

3 Once out of the water, put the pump back on and reconnect the infusion set.

4 Check your infusion set tape and the sensor tape to make sure that they are not damaged.

Utilities

Alarm review

You can review alarms and their details in the ALARM HISTORY screen. This screen shows up to 36 past alarms, errors, or LOW RESERVOIR and LOW BATTERY alerts.

1 Go to the ALARM HISTORY screen.

Main > Utilities > Alarm > Alarm History

2 Scroll through your past alarms.

3 In the ALARM HISTORY screen, select the alarm you want to review and press ACT. The details for that

alarm will appear on the screen.

4 Press ESC to return to the ALARM HISTORY screen. Select another alarm to review, or exit the menus if

you are done.

Setting your alert type

You can select the type of alert your pump uses (for alarms, special conditions and programming). You can select a vibrate (silent) alert, or an audible beep alert. There are three beep types: long, medium and short tones. The factory setting for this feature is beep-medium.

The vibrate alert type is disabled if you use the block feature, and vibrate must be set again once block is turned off. Vibrate uses more battery power than the beep alert type and may shorten battery life. If your alert type is set to vibrate and you get a LOW BATTERY alert, your pump will use the beep alert type instead to conserve battery power.

1 Go to the ALERT TYPE screen.

Main > Utilities > Alarm > Alert Type

2 Select your alert type and press ACT. That alert type is now active. Exit the menus.