ToF-SIMS is a surface analytical technique that uses an ion beam (primary ions) to remove small numbers of atoms and molecules from the outermost layer of a surface and analyzes those that are ionized (secondary ions). A short pulse of primary ions strikes the surface and the secondary ions produced in the sputtering process (illustrated in Figure 2.14) are extracted from the sample surface into a time-of-flight mass spectrometer.

Figure 2.14 Schematic representation of the SIMS sputtering process.

These secondary ions are dispersed in time according to their velocities (which are proportional to their mass/charge ratio (m/z)). Discrete packets of ions of differing mass are detected as a function of time at the detector. In general, the inherent sensitivity of mass spectrometry-based techniques, which is a consequence of the low background that

Primary Ion Beam

exists between secondary ions with mass:charge ratios that only differ by a fractional mass unit, gives SIMS a major advantage over other methods of analysis of thin films. ToF-SIMS is capable of detecting ions over a large mass range of 1 to 10,000 atomic mass units at a mass resolution (m/Δm)of 10,000. The technique is also capable of generating an image of lateral distributions of these secondary ions at spatial resolutions of better than 0.15 microns.

A dual beam mode of operation in ToF-SIMS was used in this project to study the lateral distribution of the thin oxide films on Ni-Cr-Mo and Ti alloys. A very low energy (1 keV) Bi3+ ion beam, with a target current in the range of ~7-8 nA, was employed in the dual beam mode to create a sputter crater, while the analysis of the secondary ions was performed with a higher energy (25 keV), finely focused, monoisotopic 69Ga+ ion beam.

This helps maintain the integrity of the oxide during analysis, allowing for the elucidation of both chemical and structural information. The main advantage of this technique is that a finely focused Ga ion gun achieves a smaller beam allowing for better lateral resolution. In addition, the heavier Bi3+ beam more quickly erodes the surface and enhances the yield

of negative secondary ions, which are the ions of interest emitted from an oxide film (e.g., NiO–). By using separate beams for erosion and analysis, each beam is specifically suited

to its task. This results in high elemental sensitivity and the ability to quickly erode through the surface to obtain the depth profile while maintaining excellent depth resolution, and allows for ultra shallow depth profiling within a passive oxide film to be performed on the nanometer scale [24, 25].

One disadvantage is that the depth scale cannot be quantified due to matrix effects. In a complex structure, such as a Ni-Cr-Mo oxide in which multiple components coexist,

there is a rapid succession of different matrices within a profile. This makes it difficult to accurately determine the depth resolution and depth scale for all components individually, since the sputter rates vary depending on the matrix being analyzed. Also, since standards to match the complex oxide matrix on an industrially made alloy cannot be obtained, the secondary ion intensities cannot be calibrated with any oxide mixtures of known composition. Therefore, the changes of intensity observed as a function of sputtering time in this project should be considered only qualitative.

Another disadvantage of this technique is that, while the low beam current allows for the elucidation of the chemical nature of species in the oxide, more specific characteristics, such as the oxidation state of an ion, cannot be determined. For instance, while a high intensity of CrO3– clusters detected by the mass spectrometer indicates that Cr was

bonded to oxygen in the film, it does not indicate that the oxidation state was +6. Other techniques, such as XPS, are more suited for determining oxidation states of atoms within the oxide film.

Dynamic SIMS is another mode of operation. Operated in dynamic mode, the sputtering phenomenon is the same as for static SIMS (ToF-SIMS). However, the primary beam is high enough in energy (~1 μA•cm-2) to result in a drastic change of the chemical

composition and structure of the surface. This technique is inherently “destructive” due to the collisional interaction of the primary ions with the sample, although the extent of the surface damage can be controlled by an appropriate optimization of the parameters of the primary ion source. The dynamic SIMS (D-SIMS) mode is characterized by the use of high density, DC primary ion beam currents providing typical sputter rates in the range 0.5-5 nm•s–1. Under such conditions most chemical bonds are broken and only atoms or

polyatomic fragments are ejected from the sample surface as neutrals or ions. This restricts D-SIMS applications to elemental and isotopic depth profiling or mapping. The detection range of this technique extends from matrix element levels down to trace element levels in the ppb range. It was only used in this project to map the locations of deuterium and iron in a corroded Ti-2 surface.