• No se han encontrado resultados

JUZGADO SEGUNDO MERCANTIL DEL PRIMER DEPARTAMENTO JUDICIAL DEL ESTADO

In summary, FLUKA can be used as a tool for simulating scatter. Although the program itself produces an ideal situation, the ability to add multiple geometries and materials into the setup means that a more realistic situation can be modelled.

Producing a spectrum via SpekCalc that is similar to the experimental beam is difficult to achieve. However, with some manipulation of the intensity a spectrum with sufficient simi- larities can be formed.

Using the scatter spectra attained using FLUKA and simply subtracting them from experi- mental spectra has shown a 15% increase in the contrast. This was not achievable without the spectroscopic capabilities of the HEXITEC detector and shows an alternative method of scatter removal by utilising this ability.

This improvement is less than was found when imaging using monochromatic X-rays and windowing the scatter from the spectrum. It is possible that this smaller improvement in contrast is due to the underestimation of the scatter when modelling it in FLUKA. This is the limitation of such a method as scatter is produced from surrounding objects in the room and as this is not constant for all mammographic facilities it is difficult to model an accurate scatter spectrum for all cases.

However although precise spectra of the scatter would be difficult to model, estimates can be made and have shown to have a good effect on the contrast once removed from the ex- perimental spectrum. In practise it would be possible to provide estimates of the scatter for a series of breast thicknesses which is a very simple method to introduce into the screening process without the need for expensive equipment as many methods for the improvement of image quality require.

5

Conclusions

The purpose of this study was to investigate alternative methods to the anti-scatter grid for the removal of scatter from X-ray images in mammography. Alternatives are needed because of the increase in dose that is required to compensate for the loss of statistics experienced when using an anti-scatter grid. Other modalities of breast imaging, such as breast CT and MRI, have been developed to improve on image quality but these are often more time consuming or expensive and therefore not efficient enough for the screening of women across the UK and only used for women at high risk of developing cancer or women already diagnosed. Two methods of removing scatter using the spectroscopic capabilities of a pixellated spec- troscopic detector were investigated. These methods have the advantage that they are modi- fications to conventional mammography X-ray, resulting in a practical solution to the image quality in mammography without needing very large and expensive equipment such as MRI and CT scanners and would therefore be more feasible for the use in breast screening. Both of the methods rely on the use of the pixellated spectroscopic detector HEXITEC. The detector used has a couple of limitations at this stage in its development. Firstly the detector size as the detector used in the study is only 2 x 2 cm and therefore to produce an image the size of a conventional mammogram a detector with a much larger area is required. There have been investigations made into a larger area HEXITEC detector which is 10 x 10 cm in size [99]. Although this is still not large enough for a conventional mammogram, it does show the possible development of this type of detector to a much larger size that could be used for such a purpose.

There is also a limitation in the rates that it operates at as HEXITEC is designed for low rates and imaging for a medical purpose involves high rates. The Medipix detector described in in the introduction chapter is an example of a spectroscopic detector designed for high rates but it has the limitation of only being able to access the spectroscopic information in bands instead of the full spectroscopic information acquired with HEXITEC. Therefore a possible compromise between the two detectors that dealt with higher rates for imaging with but compromised with a lesser energy resolution. The methods described both require the use of the detected energy spectra, however both methods could work equally as well with a

reduced resolution. The monochromatic method needs an energy resolution around the size of the monochromatic peak so that it can be distinguished from the scatter and the FLUKA simulations only require the full energy spectrum that scatter can be subtracted from so a lesser energy resolution could still be used.

The first method investigated was based on monochromatic X-rays produced with a con- ventional X-ray tube and a mosaic crystal monochromator. Compton-scattered X-rays from the monochromatic beam would appear at a lower energy than the primary beam in the detected spectrum and were therefore simply windowed out from the spectrum. Imaging with synchrotron monochromatic beams has been previously investigated and has seen some promising results. The downside with this method is that synchrotrons are large and expen- sive which is why a conventional X-ray tube was chosen for this study. The method used in this study also has the additional advantage of combining it with a spectroscopic detector so that scattered X-rays can be removed from the image which was not achievable before without the use of a spectroscopic detector.

Results showed an increase in contrast of 40% in comparison to imaging with a full energy beam. The main problem with this method is the limited statistics and hence the long acqui- sition time. This time was increased further as the size of the mononchromatic beam incident on the detector was very small and so multiple slices of the image were acquired and then recombined. This problem has been investigated previously by using multiple crystals to produce a much wider beam so less slices need to be acquired at one time and this is some- thing that could be investigated in order to see how much of a decrease in acquisition time it has.

The second method that was investigated was the removal of scattered X-rays by first mod- elling them using a Monte Carlo modelling package. A scatter-only spectrum was generated from an input beam with a peak energy the same as the beam used to acquire the experimen- tal image. Then using the spectroscopic capabilities of the detector, the scatter spectrum was subtracted from the experimental one; leaving a scatter-free image.

Results showed that, by subtracting the scatter from background and detail regions, the con- trast improved by about 15% when compared to the full spectrum image. The limitations

of this approach is the accuracy in modelling the scattered spectrum for every experimental situation. Apart from that the method is reasonably simple as there is limited change to a conventional mammographic set up and simulations of varying breast thicknesses could be carried out off line and the resulting scatter spectra stored. Then the scatter removal could be part of the post-processing of the image.

Further work on this method is to find a way to measure how accurate this method is from removing scatter. Then a more clinically realistic phantom could be imaged with the spec- troscopic detector and the scatter removed from the whole image.

In conclusion, this project aimed to investigate methods of removing scatter in order to im- prove image quality using a pixellated spectroscopic detector. Two methods were researched and both showed positive results by improving on the contrast of low contrast details when compared to conventional imaging.

References

[1] Advisory Committee on Breast Cancer Screening et al. Screening for breast cancer in england: past and future. Journal of Medical Screening, 13(2):59–61, 2006.

[2] Independent UK Panel on Breast Cancer Screening et al. The benefits and harms of breast cancer screening: an independent review. The Lancet, 380(9855):1778–1786, 2012.

[3] J.R. Beckett, C.J. Kotre, and J.S. Michaelson. Analysis of benefit: risk ratio and mortal- ity reduction for the uk breast screening programme. The British journal of radiology, 2014.

[4] D. Ikeda. Breast Imaging: The Requisites. Requisites in Radiology. Elsevier Health Sciences, 2010.

[5] E.S. De Paredes. Atlas of Mammography. LWW Doody’s all reviewed collection. Wolters Kluwer Health/Lippincott Williams & Wilkins, 2007.

[6] P. Butler, A. Mitchell, and H. Ellis. Applied Radiological Anatomy for Medical Stu- dents. Cambridge University Press, 2007.

[7] H. Ellis and V. Mahadevan. Clinical Anatomy: Applied Anatomy for Students and Junior Doctors. Wiley, 2011.

[8] H.L. Ioachim and L.J. Medeiros. Ioachim’s Lymph Node Pathology. LWW medical book collection. Wolters Kluwer Health/Lippincott Williams & Wilkins, 2009.

[9] J. Beutel, H.L. Kundel, and R.L. Van Metter. Handbook of Medical Imaging: Physics and psychophysics. Handbook of Medical Imaging. SPIE Press, 2000.

[10] R.E. Pollock. Advanced Therapy in Surgical Oncology. Pmph USA Ltd Series. BC Decker Incorporated, 2008.

[11] T.C. Telger and M. Lanyi. Mammography: Diagnosis and Pathological Analysis. Springer Berlin Heidelberg, 2012.

[12] D.J. Winchester and D.P. Winchester. Breast Cancer. American Cancer Society atlas of clinical oncology. B.C. Decker, 2006.

[13] A. Evans. Breast Calcification: A Diagnostic Manual. Greenwich Medical Media, 2002.

[14] J. Hodler, G.K. von Schulthess, and C. Zollikofer. Diseases of the Heart, Chest & Breast 2011-2014: Diagnostic Imaging and Interventional Techniques. Springer, 2011.

[15] M. Dustler, I. Andersson, H. Brorson, P. Fr¨ojd, S. Mattsson, A. Tingberg, S. Zackrisson, and D. F¨ornvik. Breast compression in mammography: pressure distribution patterns. Acta Radiologica, 53(9):973–980, 2012.

[16] W.R. Hendee and E.R. Ritenour. Medical Imaging Physics. Wiley, 2003.

[17] J.T. Bushberg, J.A. Seibert, E.M. Leidholdt, and J.M. Boone. The Essential Physics of Medical Imaging. Wolters Kluwer Health, 2011.

[18] P. Suetens. Fundamentals of Medical Imaging. Cambridge medicine. Cambridge Uni- versity Press, 2009.

[19] P. Allisy-Roberts and J.R. Williams. Farr’s Physics for Medical Imaging. Saunders, 2007.

[20] K.M. Mudry, R. Plonsey, and J.D. Bronzino. Biomedical Imaging. Principles and Applications in Engineering. CRC Press, 2003.

[21] R. Highnam and M. Brady. Mammographic Image Analysis. Computational imaging and vision. Kluwer Academic Publishers, 1999.

[22] C. Tang, E. Stier, K. Fischer, and H. Guckel. Anti-scattering x-ray grid. Microsystem technologies, 4(4):187–192, 1998.

[23] N.B. Smith and A. Webb. Introduction to Medical Imaging: Physics, Engineering and Clinical Applications. Cambridge Texts in Biomedical Engineering. Cambridge University Press, 2010.

[24] K.M. Kelly, J. Dean, W.S Comulada, and S. Lee. Breast cancer detection using auto- mated whole breast ultrasound and mammography in radiographically dense breasts. European radiology, 20(3):734–742, 2010.

[25] C.K. Kuhl. Mri of breast tumors. European radiology, 10(1):46–58, 2000.

[26] G. Gennaro, A. Toledano, C. Di Maggio, E. Baldan, E. Bezzon, M. La Grassa, L. Pescarini, I. Polico, A. Proietti, A. Toffoli, et al. Digital breast tomosynthesis versus digital mammography: a clinical performance study. European radiology, 20(7):1545– 1553, 2010.

[27] I. Reiser and S. Glick. Tomosynthesis Imaging. Imaging in Medical Diagnosis and Therapy. CRC Press, 2014.

[28] C.J. Strudley, K.C. Young, P. Looney, and F.J. Gilbert. Development and experience of quality control methods for digital breast tomosynthesis systems. The British journal of radiology, page 20150324, 2015.

[29] H.J. Teertstra, C.E. Loo, M. van den Bosch, H. van Tinteren, E. Rutgers, S.H. Muller, and K. Gilhuijs. Breast tomosynthesis in clinical practice: initial results. European radiology, 20(1):16–24, 2010.

[30] T. Asaga, C. Masuzawa, A. Yoshida, and H. Matsuura. Dual-energy subtraction mam- mography. Journal of digital imaging, 8(1):70–73, 1995.

[31] A. Taibi, S. Fabbri, P. Baldelli, C. Di Maggio, G. Gennaro, M. Marziani, A. Tuffanelli, and M. Gambaccini. Dual-energy imaging in full-field digital mammography: a phan- tom study. Physics in medicine and biology, 48(13):1945, 2003.

[32] P. Baldelli, A. Bravin, C. Di Maggio, G. Gennaro, M. Gambaccini, A. Sarnelli, and A. Taibi. Evaluation of the minimum iodine concentration for contrast enhanced sub- traction mammography. Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 580(2):1115– 1118, 2007.

[33] G.J. Whitman and T.M. Haygood. Digital Mammography: A Practical Approach. Cambridge medicine. Cambridge University Press, 2012.

[34] M.D. Chuntonov and M. Aslund. Physical evaluation of a prototype for the sectra microdose mammography system. In Digital Mammography: IWDM 2002—6th Inter- national Workshop on Digital Mammography, page 12. Springer Science & Business Media, 2012.

[35] P. Monnin, D. Gutierrez, S. Bulling, D. Guntern, and F.R. Verdun. A comparison of the performance of digital mammography systems. Medical physics, 34(3):906–914, 2007.

[36] S.J. Thunberg, T. Francke, J. Egerstrom, M. Eklund, L. Ericsson, T. Kristoffersson, V.N. Peskov, J. Rantanen, S. Sokolov, P. Svedenhag, et al. Evaluation of a photon- counting mammography system. In Medical Imaging 2002, pages 202–208. Interna- tional Society for Optics and Photonics, 2002.

[37] XCounter. http://www.xcounter.se/index.php?page=0.

[38] Medipix developed at CERN. http://medipix.web.cern.ch/medipix/.

[39] J.M. Boone, T.R. Nelson, K.K Lindfors, and J.A. Seibert. Dedicated breast ct: Radia- tion dose and image quality evaluation 1. Radiology, 221(3):657–667, 2001.

[40] M.K. Markey. Physics of Mammographic Imaging. Imaging in Medical Diagnosis and Therapy. CRC Press, 2012.

[41] N.D. Prionas, K.K. Lindfors, S. Ray, S. Huang, L.A. Beckett, W.L. Monsky, and J.M. Boone. Contrast-enhanced dedicated breast ct: Initial clinical experience 1. Radiology, 256(3):714–723, 2010.

[42] S.H. Heywang-Koebrunner, I. Schreer, and S. Barter. Diagnostic Breast Imaging: Mammography, Sonography, Magnetic Resonance Imaging, and Interventional Pro- cedures. Thieme, 2014.

[43] M. Nothacker, V. Duda, M. Hahn, M. Warm, F. Degenhardt, H. Madjar, S. Weinbrenner, and U. Albert. Early detection of breast cancer: benefits and risks of supplemental breast ultrasound in asymptomatic women with mammographically dense breast tissue. a systematic review. BMC cancer, 9(1):335, 2009.

[44] V. Corsetti, N. Houssami, M. Ghirardi, A. Ferrari, M. Speziani, S. Bellarosa, G. Remida, C. Gasparotti, E. Galligioni, and S. Ciatto. Evidence of the effect of ad- junct ultrasound screening in women with mammography-negative dense breasts: In- terval breast cancers at 1year follow-up. European journal of cancer, 47(7):1021–1026, 2011.

[45] M. Darboux and J. Dinten. Physical-model-based scatter correction in mammography, 1997.

[46] A. Wang, E. Shapiro, S. Yoon, A. Ganguly, C. Proano, R. Colbeth, E. Lehto, and J. Star-Lack. Asymmetric scatter kernels for software-based scatter correction of grid- less mammography, 2015.

[47] J.L. Ducote and S. Molloi. Scatter correction in digital mammography based on image deconvolution. Physics in medicine and biology, 55(5):1295, 2010.

[48] T.W. Freer and M.J. Ulissey. Screening mammography with computer-aided detec- tion: Prospective study of 12,860 patients in a community breast center 1. Radiology, 220(3):781–786, 2001.

[49] J. Tang, R.M. Rangayyan, J. Xu, I. El Naqa, and Y. Yang. Computer-aided detec- tion and diagnosis of breast cancer with mammography: recent advances. Information Technology in Biomedicine, IEEE Transactions on, 13(2):236–251, 2009.

[50] C.M. Ronckers, C.A. Erdmann, and C.E. Land. Radiation and breast cancer: a review of current evidence. Breast Cancer Res, 7(1):21–32, 2005.

[51] C. Suryanarayana and G. Norton. X-Ray Diffraction: A Practical Approach. Springer US, 2013.

[52] S.N. Ahmed. Physics and Engineering of Radiation Detection. Elsevier Science, 2014. [53] A. Sarnelli, A. Taibi, A. Tuffanelli, G. Baldazzi, D. Bollini, A.E. Cabal Rodriguez, M. Gombia, F. Prino, L. Ramello, E. Tomassi, et al. K-edge digital subtraction imaging based on a dichromatic and compact x-ray source. Physics in medicine and biology, 49(14):3291, 2004.

[54] R. Lewis. Medical applications of synchrotron radiation x-rays. Physics in medicine and biology, 42(7):1213, 1997.

[55] A.K. Freund, A. Munkholm, and S. Brennan. X-ray diffraction properties of highly oriented pyrolytic graphite. In SPIE’s 1996 International Symposium on Optical Sci- ence, Engineering, and Instrumentation, pages 68–79. International Society for Optics and Photonics, 1996.

[56] M.S. del Rio, M. Gambaccini, G. Pareschi, A. Taibi, A. Tuffanelli, and A.K. Freund. Focusing properties of mosaic crystals. In SPIE’s International Symposium on Optical Science, Engineering, and Instrumentation, pages 246–255. International Society for Optics and Photonics, 1998.

[57] F. Arfelli, A. Bravin, G. Barbiellini, G. Cantatore, E. Castelli, M. Di Michiel, P. Poropat, R. Rosei, M. Sessa, A. Vacchi, et al. Digital mammography with syn- chrotron radiation. Review of scientific instruments, 66(2):1325–1328, 1995.

[58] E. Burattini, E. Cossu, C. Di Maggio, M. Gambaccini, P.L. Indovina, M. Marziani, M. Pocek, S. Simeoni, and G. Simonetti. Mammography with synchrotron radiation. Radiology, 195(1):239–244, 1995.

[59] R. Moeckli, F.R. Verdun, S. Fiedler, M. Pachoud, P. Schnyder, and J. Valley. Objective comparison of image quality and dose between conventional and synchrotron radiation mammography. Physics in medicine and biology, 45(12):3509, 2000.

[60] E. Burattini, M. Gambaccini, M. Marziani, O. Rimondi, P.L. Indovina, M. Pocek, G. Simonetti, M. Benassi, C. Tirelli, and R. Passariello. X-ray mammography with synchrotron radiation. Review of scientific instruments, 63(1):638–640, 1992.

[61] R. Longo, M. Tonutti, L. Rigon, F. Arfelli, D. Dreossi, E. Quai, F. Zanconati, E. Castelli, G. Tromba, and M.A. Cova. Clinical study in phase-contrast mammog- raphy: image-quality analysis. Philosophical Transactions of the Royal Society of London A: Mathematical, Physical and Engineering Sciences, 372(2010):20130025, 2014.

[62] E. Castelli, M. Tonutti, F. Arfelli, R. Longo, E. Quaia, L. Rigon, D. Sanabor, F. Zan- conati, D. Dreossi, A. Abrami, et al. Mammography with synchrotron radiation: first clinical experience with phase-detection technique. Radiology, 259(3):684–694, 2011.

[63] P. Baldelli, A. Taibi, A. Tuffanelli, and M. Gambaccini. Quasi-monochromatic x-rays for diagnostic radiology. Physics in medicine and biology, 48(22):3653, 2003.

[64] P. Baldelli, A. Taibi, A. Tuffanelli, M.C. Gilardoni, and M. Gambaccini. A prototype of a quasi-monochromatic system for mammography applications. Physics in medicine and biology, 50(10):2225, 2005.

[65] M. Gambaccini, A. Taibi, A. Del Guerra, F. Frontera, and M. Marziani. Narrow energy band x-rays via mosaic crystal for mammography application. Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 365(1):248–254, 1995.

[66] M. Gambaccini, A. Tuffanelli, A. Taibi, A. Fantini, and A. Del Guerra. Bragg- diffraction-based quasi-monochromatic source for mammography using mosaic crys- tals. In SPIE’s International Symposium on Optical Science, Engineering, and Instru- mentation, pages 174–184. International Society for Optics and Photonics, 1999.

[67] D. Vavrik, J. Dammer, J. Jakubek, I. Jeon, O. Jirousek, M. Kroupa, and P. Zlamal. Ad- vanced x-ray radiography and tomography in several engineering applications. Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 633:S152–S155, 2011.

[68] L.S. Pinsky, J. Idarraga-Munoz, M. Kroupa, H.M. Son, N.N. Stoffle, E.J. Semones, A.A. Bahadori, D. Turecek, S. Posp´ıˇsil, J. Jakubek, et al. Medipix in space on-board the iss. Journal of radiation research, 55(suppl 1):i62–i63, 2014.

[69] R. Van Gastel, I. Sikharulidze, S. Schramm, J.P. Abrahams, B. Poelsema, R.M. Tromp, and S.J. Van Der Molen. Medipix 2 detector applied to low energy electron microscopy. Ultramicroscopy, 110(1):33–35, 2009.

[70] T.R. Melzer, N.J. Cook, A.P. Butler, R. Watts, N. Anderson, R. Tipples, and P.H. Butler. Spectroscopic biomedical imaging with the medipix2 detector. Australasian Physics & Engineering Sciences in Medicine, 31(4):300–306, 2008.

[71] L. Jones, P. Seller, M. Wilson, and A. Hardie. Hexitec asic—a pixellated readout chip for czt detectors. Nuclear Instruments and Methods in Physics Research Section A: Ac- celerators, Spectrometers, Detectors and Associated Equipment, 604(1):34–37, 2009.

[72] S.D. Christe, A. Shih, M. Rodriguez, A. Cramer, K. Gregory, M. Edgerton, J. Gaskin, C. Wilson-Hodge, J. Apple, K. Chavis, et al. The high energy replicated optics to explore the sun mission: a hard x-ray balloon-borne telescope. In SPIE Optical En- gineering+ Applications, pages 886206–886206. International Society for Optics and Photonics, 2013.

[73] C. Christodoulou, C.B. Reid, D. O’Flynn, M. Wilson, M. Veale, R.J. Cernik, P. Seller, and R.D. Speller. Multivariate analysis of pixelated diffraction data. Journal of Instru- mentation, 6(12):C12027, 2011.

[74] J.W. Scuffbam, M.D. Wilson, S. Pani, D.D. Duarte, M.C. Veale, S. Bell, P. Seller, P.J. Sellin, and R.J. Cernik. Evaluation of a new small-pixel cdte spectroscopic detector in dual-tracer spect brain imaging. In Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), 2012 IEEE, pages 3115–3118. IEEE, 2012.

[75] S. Pani, S.C. Saifuddin, C. Christodoulou, M. Veale, P. Seller, R.D. Speller, M. Wilson, and J.W. Scuffham. K-edge subtraction imaging using a pixellated energy-resolving

detector. In SPIE Medical Imaging, pages 79614C–79614C. International Society for Optics and Photonics, 2011.

[76] U. Bick and F. Diekmann. Digital Mammography. Medical Radiology. Springer, 2010.

[77] M.A. Flower. Webb’s Physics of Medical Imaging, Second Edition. Series in Medical Physics and Biomedical Engineering. Taylor & Francis, 2012.

[78] S. Smith. Digital Signal Processing: A Practical Guide for Engineers and Scientists: A Practical Guide for Engineers and Scientists. Demystifying technology series. Elsevier Science, 2013.

[79] G.D. Boreman. Modulation Transfer Function in Optical and Electro-optical Systems. SPIE tutorial texts. Society of Photo Optical, 2001.

[80] D.R. Dance, C.L. Skinner, K.C. Young, J.R. Beckett, and C.J. Kotre. Additional factors for the estimation of mean glandular breast dose using the uk mammography dosimetry protocol. Physics in medicine and biology, 45(11):3225, 2000.

[81] Guidelines on quality assurance visits. Sheffield, NHS Breast Screening Program, NHS- BSP Publication No 40:51–52, 1998.

[82] M.D. Wilson, S.J. Bell, R.J. Cernik, C. Christodoulou, C.K. Egan, D. O’Flynn, S. Jacques, S. Pani, J. Scuffham, P. Seller, et al. Multiple module pixellated cdte spec- troscopic x-ray detector. IEEE Trans. Nucl. Sci., 60:1197–1200, 2013.

[83] J.A. Bearden. X ray wavelengths. NYO-10586. U.S. Atomic Energy Commission, Division of Technical Information Extension, 1964.

[84] S. Pani. Private communication.

[85] G.F. Knoll. Radiation Detection and Measurement. Wiley, 2000.

[86] M.C. Veale, J. Kalliopuska, H. Pohjonen, H. Andersson, S. Nenonen, P. Seller, and M.D. Wilson. Characterization of m-π-n cdte pixel detectors coupled to hexitec readout chip. Journal of Instrumentation, 7(01):C01035, 2012.

[87] Spekcalc. http://spekcalc.weebly.com.

[88] Gammex. http://www.gammex.com/.

[89] L.M. Webb and D.A. Auger. Use of flat-field correction and adaptive linear interpo- lation to improve mammography image quality. In Proceedings of the 22nd Annual Symposium of the Pattern Recognition Association of South Africa, pages 150–155, 2011.

[90] NIST X ray transition energies database. http://www.nist.gov/pml/data/xraytrans/.

[91] K.C. Young, A. Burch, and J.M. Oduko. Radiation doses received in the uk breast screening programme in 2001 and 2002. The British journal of radiology, 2014.

[92] Newport ILS linear motors. http://www.newport.com/.

[93] H. Aichinger, J. Dierker, S. Joite-Barfuß, and M. S¨abel. Radiation Exposure and Image Quality in X-Ray Diagnostic Radiology: Physical Principles and Clinical Applications. Springer Berlin Heidelberg, 2013.

[94] L. Shao, R. Freifelder, and J.S. Karp. Triple energy window scatter correction technique in pet. Medical Imaging, IEEE Transactions on, 13(4):641–648, 1994.

[95] A.L. Kwan, J.M. Boone, and N. Shah. Evaluation of x-ray scatter properties in a dedi- cated cone-beam breast ct scanner. Medical physics, 32(9):2967–2975, 2005.

[96] FLUKA monte carlo simulation package. http://www.fluka.org/fluka.php.

[97] A. Ferrari, P.R. Sala, A. Fasso, and J. Ranft. Fluka: A multi-particle transport code