• No se han encontrado resultados

6. Beneficio del plan de gestión de seguridad y salud ocupacional

6.5. Índices de seguridad del proyecto

6.5.5. Programa anual de capacitaciones

As was mentioned in section 3.2, the first step to couple electromagnetically

an EMAT to a conducting sample is injecting an alternating current through its

coil. The current density induced in the sample is concentrated in the upper surface

and decreases exponentially with depth. The skin depth or depth of penetration,

by definition is the depth at which the current density is attenuated by 1/e(37% or -8.7 dB)[14], and depends on the frequency of the current, the properties of the

conducting sample, its conductivity or resistivity and its permeability, and coil ge-

ometry and lift-off.

In order to compute and plot this effect, and to quantify the current density induced

by the spiral coil of the PE-EMAT system presented in this body of work, a model

was implemented in Comsol Multiphysics 3.5a, using the 2D-Axial symmetry mode

available in the AC/DC module with the time-harmonic formulation; which assumes

Material Conductivity (σ [S/m]) Relative permeability (µr)

Aluminium 3.802x107 1

Steel 5.882x106 100

Copper 5.998x107 1

Table 5.1: Properties of the materials employed in the model implemented in Comsol to compute and plot the current density induced by the spiral coil of the PE-EMAT system.

Due to the symmetry of the situation, just half of the coil is necessary in the model.

Therefore, the coil has a 6.1 mm radius, each turn in the coil (18 turns in total) has

a 0.3 mm diameter, and is separated from the sample by 0.25 mm (lift-off due to

the protective face of the coil). The sample is represented as a cylinder of radius 7

mm and height 0.18 mm. The depth was selected to ensure that the current density

has attenuated completely, thus is more than three times the maximum skin depth

to model; which corresponds to the aluminium sample (3*d= 0.114 mm).

The sample is either aluminium or steel, and the coil is considered to be copper;

their material properties are presented in table 5.1. The current in the coil was

considered to have a magnitude of 45 A and it is assumed a single frequency of 4.5

MHz; which is selected because the frequency spectrum of the generation current

measured experimentally is centred around that frequency (see section 5.3). The

schematic diagram of the model used in the simulation, and the 3D geometry ob-

tained by revolving (90o) the 2D geometry about the z-axis, are shown in figure

5.29.

The results obtained from the simulation in both samples are shown in figure 5.30.

As can be seen, in both cases the current density decreases exponentially, as theory

predicts. For the steel sample, the magnitude of the current density at the surface

is 4.66x107 Am−2, and at a depth equal to the skin depth (d= 0.98x10−5 m) has

(a) 2D-axial geometry

(b) 3D geometry

Figure 5.29: Schematic diagram of the model used in the simulation of the induced current density by the EMAT coil in different samples, and the 3D geometry obtained by revolving 90o the 2D geometry about the z-axis.

the aluminium sample the current density at the surface is 3.57x107 Am−2, and at a

depth equal to the skin depth (d= 0.38x10−4 m), has decreased to 1.26x107 Am−2;

which represents a reduction of 35.7%.

Another interesting feature that can be obtained from the simulation is the varia-

tion of the dynamic field when the gap between the coil and the sample is increased.

As was mentioned in section 5.9, the ultrasonic signal amplitude will decrease with

increased lift-off, as both the bias and dynamic fields at the surface of the sample

will decrease. In the previous section it was shown how the bias field would decrease

when the gap between the sample and the electromagnet has been increased (891

0 0.5 1 1.5 2 x 10−5 0 2 4 6 8 x 10−3 0 1 2 3 4 5 x 107 Radial position /m EMAT coil (18 turns) on Steel sample

Depth into sample /m

Magnitude of the current density /Am

−2 @ Skin depth (~0.98x10−5 m) 1.66x107 Am−2 @ sample surface 4.66x107 Am−2 (a) Steel 0 0.2 0.4 0.6 0.8 1 x 10−4 0 2 4 6 8 x 10−3 0 1 2 3 4 x 107 Radial position /m EMAT coil (18 turns) on Aluminium sample

Depth into sample /m

Magnitude of the current density /Am

−2 @ sample surface 3.57x107 Am−2 @ Skin depth (~0.38x10−4 m) 1.26x107 Am−2 (b) Aluminium

Figure 5.30: Induced current density decay in steel and aluminium samples by the generation current flowing in the EMAT coil; which were obtained from the simulation in Comsol. Note the order of magnitude difference on the “Depth into sample” axes on the above plots.

the dynamic field obtained from the simulations in Comsol is presented. In a simi-

lar fashion as with the bias field, the magnitude of the dynamic field decreases with

lift-off. Another interesting feature that is shown in figure 5.31, is that increasing

the lift-off would have the effect of making a point in the coil more sensitive between

the centre and the edge. For instance, the profile of the dynamic field for a lift-off

of 2 mm shows that the coil would be more sensitive between a radial position of

3-4 mm.

In absolute terms, the coil will be less sensitive for increased lift-off, and it is worth

noting that the coil will always have some degree of lif-off via a protection face;

which can be an epoxy resin layer for EMAT coils working at room temperature, or

a ceramic plate for those working at higher temperatures.

40 30 20 10 0 Dynamic Field (B c ) /mT 8 6 4 2 0 Radial position /mm Lift-off 0 mm 0.5 mm 1.0 mm 1.5 mm 2.0 mm

Figure 5.31: Dynamic magnetic field profiles across the EMAT coil radius for a

number of different lift-off positions, when considering the steel sample. The results

Documento similar