• No se han encontrado resultados

en los Fines de la Entidad o Disolución de la Misma

In document CÓDIGO ADMINISTRATIVO (página 157-0)

The sample preparation for BSE-SEM was more laborious. First the specimen needed to be broken

and placed in a small plastic container, approximately 1.5 cm in diameter. Then in a cup enough

resin to cover the specimen was made at a ratio of 25:3 g of EpoFix Resin to EpoFix Hardener

(Figure 30a). The resin was introduced into the container under a vacuum to ensure the minimum

amount of air bubbles got stuck with the resin (Figure 30b). Once the resin completely covered

the specimen, the specimen was put aside, and the resin was allowed to harden over a period of 24

hours (Figure 30c). A Struers Accutom-50 diamond saw was used to cut the specimen such that

it had a planar surface (Figure 30d-e). A Struers grinding polishing machine was then used to

polish the specimen through a series of decreasing grain size (Figure 30f-h). After polishing the

specimen was set aside for another 24 hours to allow enough time for the polishing alcohol residue

to evaporate (Figure 30i). The specimen was then coated with a carbon coater (Leica) to create a

conductive surface on the plane (Figure 30c).

(a) (b)

54

(c) (d)

(e) (f)

55

(i) (j)

Figure 30 Sample Preparation for back scattering electron scanning electron microscopy.

4.5. Mercury Intrusion Porosimetry

Mercury intrusion porosimetry (MIP) was used to characterize the pore structure of the CNF-

UHPC composites. This technique is based on the intrusion of a non-wetting fluid (mercury) into

the connected pore structure under increasing pressure. The Washburn equation is used to relate

the pressure to the entry pore size. Samples were crushed to obtain a mass of 1g. The dried samples

were placed in the dilatometer and the air was removed. The measurement was in two steps for

two different populations of the pores. In the first step a pressure of 100kPa was applied to intrude

the larger pores. In the second step, the pressure was increased up to 400 MPa which allowed the

mercury to intrude pore entries down to 2 nm.

4.6. Inverse Gas Chromatography

A Hewlett-Packard gas chromatograph fitted by Surface Measurement Systems (SMS) with a

flame ionization detector (FID) was used for the Inverse Gas Chromatography (IGC)

measurements. Glass columns of 30 cm in length and 4 cm in diameter were packed with a tap-

and-fill method with specimens that were grounded up by hand and then sieved through a #200 or

#140 sieve. Each column was packed with 100 mg – 150 mg of material. Before the columns

56

were at 60°C with an injection flow rate of 10 sccm. The probes listed in Table 9 were selected

based off the limited literature on IGC and cement [146,162-164]. The dead-time was determined

by methane injection. A Flame Ionization Detector (FID) was used to determine retention times.

Table 9 Selected probes used in this study.

Probe Molecule Type Probe Molecule Type

Benzene aromatic Methane n-alkane

Chloroform aromatic Octane n-alkane

Ethyl acetate polar Nonane n-alkane

Methanol polar Decane n-alkane

Acetonitrile polar Undecane n-alkane

4.7. Dynamic Light Scattering

The concentration of particles remaining in the filtered interstitial pore fluid after mixing and

filtering was determined by Dynamic Light Scattering (DLS). Measurements were performed

using a Zetasizer nano S from Malvern Instruments, operating at an incident light wavelength of

633 nm, scattered light detection angle of 173°, and constant temperature of 25 °C. The value of

the scattered intensity was computed from the average of five independent measurements. DLS is

generally used to measure the size distribution by intensity of particles or polymer coils in a liquid

medium. In this work, the measured scattered intensity was further used to study the nanoparticles

in cement pore solution and estimate their concentration.

Plotted in Figure 31a, the size distribution by intensity of cement pore solutions solution at

different relative concentrations obtained through calcium hydroxide dilution. The calcium

hydroxide solution (i.e. lime) was prepared by adding to distilled water 0.85g/L of Ca(OH)2 . The

presence of nanoparticles with a size around 130nm can be seen. Figure 31b shows that the light

intensity scattered by these nanoparticles, which corresponds to the peak area from 30 nm to 1000

57

work, the concentration of nanoparticles in the pore solution will be estimated by measuring the

scattered intensity (i.e. peak area).

Figure 31 (a) The size distribution by intensity of cement pore solutions at different relative

concentrations (i.e. the ratio between concentration of the solution after dilution with calcium

hydroxide solution and the concentration of the initial solution) and (b) the corresponding peak

area as a function of the relative concentration. The pore solution is obtained from a cement

paste prepared at W/C=0.3, containing 0.4% PCE and mixed at the speed of 840 rpm.

4.8. Adsorption Isothermals

Total Organic Carbon (TOC) analysis was used to measure the adsorption of the PCE on the

surfaces of the cement or quartz powders. The TOC analyzer in this study was manufactured by

Shimadzu. The adsorption levels of different dosages of PCE on the mineral phases of the powder

were found by taking the difference between the total organic carbon content of a reference PCE

solution and the amount of PCE extracted from the pore solution of a cement or quartz powder

paste. The amount of organic carbon content in the powders from the production process cannot 0.0E+00 2.0E+04 4.0E+04 6.0E+04 8.0E+04 1.0E+05 1.2E+05 1.4E+05 1.6E+05 10 100 1000 In te n sity (k cp s) Radius (nm) Relative concentration =0.4 Relative concentration =0.5 Relative concentration =1 y = 715720x R² = 0.99 0.0E+00 1.0E+05 2.0E+05 3.0E+05 4.0E+05 5.0E+05 6.0E+05 7.0E+05 8.0E+05 0.20 0.70 1.20 In te n sity (k cp s) Relative concentration (-)

58

be neglected as it strongly affects the measurements if left aside. The subsequent measurements

were taken from pastes with incremental increases in PCE dosage. The adsorption isotherms were

then computed to determine the saturation levels of the PCE on the surfaces of the cement or quartz

powders.

4.9. Isothermal Calorimetry

The thermal power and heat of hydration of the cement pastes were monitored for the first 48 hours

using a TAM Air microcalorimeter (Thermometrics) at a constant temperature of 20 ºC. Samples

were mixed according to the mixing procedure and then 5g of paste was placed inside the

calorimeter cell. All samples were balanced with a reference cell having the same heat capacity.

In document CÓDIGO ADMINISTRATIVO (página 157-0)